Users Manual

80
Registration and selection of the DR Detector
3�4�2 Registration with AeroDR
Battery Charger2
1
Make sure that all the devices in the desti-
nation exposure room are usable�
2
Insert the DR Detector into AeroDR Bat-
tery Charger2 in the destination exposure
room�
Registration process will start.
Exposure
side
3
Oncetheregistrationiscompleted,a
buzzerwillsoundontheAeroDRBattery
Charger2,andtheLED(green)willchange
fromslowashingtolit.
LED (green)
4
ConrmthattheDRDetectoriconisdis-
played on the image processing control-
ler
HINT
Charging will also start when the DR Detector is inserted
into the AeroDR Battery Charger2, and the AeroDR Battery
Charger2 LED (blue) will light or ash.
3�4�3 Registration with the wired
cable
1
Make sure that all the devices in the desti-
nation exposure room are usable�
2
Securely connect the wired cable to the
wired connection connector on the DR
Detector
Registration process will start.
or
3
ConrmthattheDRDetectoriconisdis-
played on the image processing control-
ler