Users Manual

1
2
3
4
5
6
7
8
9
10
11
12
Operation of DR Detector
65
3
PushtheexposureswitchoftheX-rayde-
vice to perform the exposure�
After exposure, images are saved to DR Detec-
tor, and the information LED indicates the num-
ber of saved images.
4
Exit Aero Storage mode�
Insert DR Detector into the AeroDR Battery
Charger or AeroDR Battery Charger2, or con-
nect it to a wired cable.
Aero Storage mode exits, and DR Detector is
connected to the image processing controller.
5
Import the images using the image pro-
cessing controller
Aero Indicator
Exposure with Aero Indicator is performed with the fol-
lowing procedure.
IMPORTANT
Do not give diagnoses using the images displayed in Aero
Indicator.
Images displayed in Aero Indicator are for positioning
check, and the display di󰀨ers from the image displayed in
the image processing controller.
When performing exposure using Aero Indicator, align the
orientation of the triangular mark of the DR Detector with
the orientation of the grid bar as shown in the following g-
ure.
DR DetectorGrid
Grid bar
Triangular mark
If the orientation of the grid bar tilts 90° from the orientation
of the triangular mark of the DR Detector at exposure, the
image display unevenness may occur on Aero Indicator.
The image unevenness on Aero Indicator due to the orien-
tation of the grid bar will be removed upon image process-
ing when the image is imported to the image processing
controller.
1
Start the Aero Storage mode of the DR De-
tector
2
Start Aero Indicator of the DR Detector to
beaccessedfromamobiledevice,and
log in to Aero Indicator
3
ChecktheconnectedDRDetector,andtap
[OK]�
4
Ensure that the status of Aero Indicator is
readyforexposure,andperformanexpo-
sure operation�
5
Check the preview image displayed in
Aero Indicator� The image is displayed
duringthereceptionprocess,andthepre-
view image is displayed�
6
Exit the Aero Storage mode of the DR
Detector as well as Aero Indicator on the
mobile terminal�
7
Import the image using the image pro-
cessing controller
HINT
If reconnect to the DR Detector that imported images to the
image processing controller without exiting Aero Indicator,
the following screen is displayed. Tap [Refresh] to refresh
the imported image as well as exposure failure and division
display displayed on the mobile terminal.