Various problems and countermeasures
Status Error description Corrective actions
No image is acquired.
*2
No image appears after X-ray exposure.
(1)Check the following items.
Did you perform exposure in "Ready" sta-
tus?
Are the exposure conditions correct?
Was X-ray output lowered?
Was X-ray voltage set to a lower value?
(2)Perform the following procedures.
For AeroDR Detector
Set the sensitivity setting of the image
processing controller to "Large" before
exposure. If the setting is already set to
"Large", increase the voltage of the X-ray
and perform exposure.
For AeroDR 2 Detector and AeroDR 3 De-
tector
Increase the voltage of the X-ray before
exposure.
Reading has automatically
started.
*2
The machine starts reading the image without
X-ray exposure.
(1)Check the following items.
For AeroDR Detector
Was the DR Detector subjected to strong
shocks or vibrations while "Ready" was
displayed?
Is the sensitivity setting of the image
processing controller correct? (Is the
sensitivity setting of the image processing
controller set to "Large"?)
For AeroDR 2 Detector and AeroDR 3 De-
tector
Was the DR Detector subjected to strong
shocks or vibrations while "Ready" was
displayed or wasn't an X-ray exposed
nearby?
(2) Use caution to prevent strong shocks or
vibrations and restart exposure.
Exposure was performed in a
condition other than when the ex-
posure with the image processing
controller was possible.
*2
Exposure was performed when "Ready" was
not displayed on the image processing con-
troller. Consequently, a correct image was not
acquired.
Wait for one minute or more before starting
the next exposure.
Exposure was performed when "Ready" was
not displayed.
Wait for 30 seconds or more before starting
the next exposure.
*1 The AeroDR 1012HQ and AeroDR 3 1012HQ cannot be inserted into the AeroDR Battery Charger.
*2 Troubleshooting in the Aero Sync mode.