Various problems and countermeasures
Corrective actions when the DR Detector is dropped or subjected to impacts
Run a test using the drop/impact panel check tool of the image processing controller.
IMPORTANT
The drop/impact panel check tool cannot detect all errors.
If you use the DR Detector after the test and nd any error in the image, contact Konica Minolta technical representatives.
Use the DR Detector as is.
PASS
FAIL
The DR Detector was dropped or
subjected to impacts.
Contact Konica Minolta technical
representatives.
Start testing with the
drop/impact panel check tool of the image
processing controller.
HINT
If a message is displayed in the image processing controller using the AeroDR 2 Detector, and AeroDR 3 Detector, follow the
message and run a test using the drop/impact panel check tool.
When the AeroDR 3 Detector detect an impact, the Information LED (white) ashes slowly. Run a test using the drop/impact
panel check tool.