Various problems and countermeasures
5�2�3 AeroDR Interface Unit2
Status Error description Corrective actions
Power LED (green) does not
light.
AC power not supplied.
Make sure the power cable is connected cor-
rectly.
Feeding1/2 LED (blue) does not
light.
Feeding1/2 LED (blue) does not light even
when connecting the DR Detector.
Make sure that the I/F Cable is connected cor-
rectly.
Communications are down be-
tween devices, and power LED
(green) is o󰀨.
The power switch is not on. Or, the power
cable is not connected.
Make sure the power switch of the AeroDR
Interface Unit2 is on. Or, make sure the power
cable is connected correctly.
The DR Detector is being used
with wired connection, but it is
not recognized on the image pro-
cessing controller, and feeding1/2
LEDs (blue) are o󰀨.
The I/F Cable is not connected.
Make sure that the I/F Cable is connected cor-
rectly.
Battery is not charging while the
DR Detector is used with wired
connection (the battery level on
the battery level display of the
image processing controller does
not increase), and feeding1/2
LEDs (blue) are o󰀨.
5�2�4 Detector Interface Unit
Status Error description Corrective actions
LED(green/blue) does not light. AC power not supplied.
Make sure that the Ethernet cable is con-
nected to the Power Supply Unit properly. Or,
make sure that the DI Unit AC adapter is con-
nected correctly.
LED (blue) does not light.
The I/F cable is not connected to the DR De-
tector.
Make sure that the DR Detector is connected
to the I/F cable correctly.
The DR Detector is being used
with wired connection, but it is
not recognized on the image
processing controller.
The I/F cable is not connected to the DR De-
tector. Or, the Ethernet cable is not connected
to the Power Supply Unit or the hub.
Make sure that the DR Detector is connected
to the I/F cable correctly. Or, make sure that
the Ethernet cable is connected correctly.