6�1 Maintenance and inspection items
This chapter describes the inspections and cleaning
required in order to maintain the use of this device in
an optimum condition.
QualityTool
HINT
“Simple Check QC for CS-7” is available for checking of im-
aging quality. Please contact our customer service for the
detail.
For details, contact Konica Minolta technical representa-
tives.
6�1�1 Maintenance schedule
The maintenance and inspection items that the user
should perform are as follows.
IMPORTANT
To ensure optimum use of this device, be sure to perform
periodic maintenance.
The following task intervals are estimates and vary accord-
ing to usage.
Maintenance task
Maintenance
interval
Checking and cleaning the surface of the
DR Detector
Weekly
Checking for external damage to the DR
Detector
Weekly
Cleaning the spring connectors of the
wired cable and AeroDR Battery Char-
ger2
Weekly
Cleaning the wired connection connec-
tors of the DR Detector and AeroDR UF
Cable
Weekly
Cleaning the AeroDR Battery Charger
and AeroDR Battery Charger2
Weekly
Full charge of the DR Detector Monthly
Calibration
Every year or
when a message
is displayed
HINT
Perform calibration when a message that prompts for cali-
bration of the image processing controller is displayed.
6�1�2 Cleaning
IMPORTANT
Be careful not to apply any cleaning chemical or liquid onto
the spring connectors, the wired connection connectors,
and the LEDs.
Do not clean with sharp or hard metal objects. If you cannot
remove stains, contact Konica Minolta technical represen-
tatives.
Wear and deformation of the protective cover will occur
with the DR Detector due to the way it is handled. The pro-
tective cover can be replaced for a fee when the damage
becomes extensive, so contact Konica Minolta technical
representatives.