Users Manual

120
Specications
Item Description
Maximum patient weight
Point load:
AeroDR 1417HQ : 150 kg@Φ40 mm
AeroDR 1417S : 150 kg@Φ40 mm
AeroDR 1717HQ : 150 kg@Φ40 mm
AeroDR 1012HQ : 150 kg@Φ40 mm
AeroDR 2 1417HQ : 150 kg@Φ40 mm
AeroDR 2 1417S : 150 kg@Φ40 mm
AeroDR 3 1417HD : 180 kg@Φ40 mm
AeroDR 3 1717HD : 180 kg@Φ40 mm
AeroDR 3 1012HQ : 180 kg@Φ40 mm
AeroDR 3 1417HD2 : 180 kg@Φ40 mm
AeroDR 3 1717HD2 : 180 kg@Φ40 mm
Face load:
AeroDR 1417HQ : 300 kg@e󰀨ective image area overall
AeroDR 1417S : 300 kg@e󰀨ective image area overall
AeroDR 1717HQ : 300 kg@e󰀨ective image area overall
AeroDR 1012HQ : 300 kg@e󰀨ective image area overall
AeroDR 2 1417HQ : 300 kg@e󰀨ective image area overall
AeroDR 2 1417S : 300 kg@e󰀨ective image area overall
AeroDR 3 1417HD : 400 kg@e󰀨ective image area overall
AeroDR 3 1717HD : 400 kg@e󰀨ective image area overall
AeroDR 3 1012HQ : 400 kg@e󰀨ective image area overall
AeroDR 3 1417HD2 : 400 kg@e󰀨ective image area overall
AeroDR 3 1717HD2 : 400 kg@e󰀨ective image area overall
* Deadweight, even when loaded on the DR Detector, has no e󰀨ect on images and the DR
Detector. The measurement method is based on KM standards.
Communication
AeroDR 1417HQ, AeroDR 1417S, AeroDR 1717HQ and AeroDR 1012HQ:
Dedicated wired Ethernet connection / wireless LAN (IEEE802.11a compliant)
AeroDR 2 1417HQ, AeroDR 2 1417S, AeroDR 3 1417HD, AeroDR 3 1717HD, AeroDR 3
1012HQ, AeroDR 3 1417HD2 and AeroDR 3 1717HD2:
Dedicated wired Ethernet connection / wireless LAN (IEEE802.11a/n compliant)
Drive power Dedicated power cable drive / battery drive
WLAN encryption
Wireless encryption method : AES
Authentication method : WPA2-PSK
Patient doses Patient doses are equivalent to 500-1000 speed lm/screen systems.
Required time for wired/wireless
switching
Less than 2 seconds
Battery type Lithium ion capacitor
Number of exposable images
AeroDR 1417HQ : 211 images/5.8 hours
AeroDR 1417S : 211 images/5.8 hours
AeroDR 1717HQ : 189 images/5.2 hours
AeroDR 1012HQ : 146 images/4.0 hours
AeroDR 2 1417HQ : 300 images/8.2 hours
AeroDR 2 1417S : 150 images/4.1 hours
AeroDR 3 1417HD : 100 [um] 251 images/6.9 hours
200 [um] 309 images/8.6 hours
AeroDR 3 1717HD : 100 [um] 217 images/6.0 hours
200 [um] 276 images/7.6 hours
AeroDR 3 1012HQ : 100 [um] 145 images/3.9 hours
200 [um] 165 images/4.5 hours
AeroDR 3 1417HD2 : 100 [um] 217 images/6.0 hours
200 [um] 272 images/7.5 hours
AeroDR 3 1717HD2 : 100 [um] 190 images/5.2 hours
200 [um] 245 images/6.8 hours
* Under conditions that the interval between studies is ve minutes and three images are
captured in each study, assuming 20 seconds for each exposure to position a patient (when
connected to the CS-7).