Safety precautions
17
CAUTION
Take the following notes when using the DR Detector:
Use the DR Detector on a at base. If the DR Detec-
tor is placed on a sloping surface and subjected to a
load, its internal sensor may be damaged.
When laying the patient onto the DR Detector during
positioning, do so very slowly.
When placing the DR Detector under the patient as
part of the exposure process, insert and pull out the
detector slowly.
Use the recommended adapter when you need to
perform exposure on a stretcher or a place where
load is to be applied locally.
Exposing or immersing the DR Detector to patient
body uids, chemicals, water, etc. may allow such liq-
uids to enter through exterior gaps and cause a fail-
ure. Make sure to cover the DR Detector with a vinyl
sheet, etc.
Be careful of the ingress of patient body uids, chem-
icals, water, etc., because the AeroDR Detector is not
designed to be waterproof.
The AeroDR 2 Detector and AeroDR 3 Detector are
designed to have waterproof capabilities, but be care-
ful when handling them because they are not guar-
anteed to be completely waterproof. Also, be careful
because subjecting them to strong impacts or exces-
sive loads by dropping or hitting them against other
objects may diminish their waterproof capabilities.
Never leave DR Detectors in hot and humid environ-
ments for long periods of time.
The DR Detector has wireless antennae positioned
at 2 places. Do not block them with the body or metal
because that will interfere with, disconnect or slow
down wireless communication.
Pressing the power switch on the AeroDR Detector
and AeroDR 2 Detector causes each LED (green,
orange, blue) to light temporarily or ash. After this,
only green begins to flash in a slow cycle. Please
check the lighting or ashing of each color once.
Pressing the Power SW of the AeroDR 3 Detector
causes the Information LED (white) to ash or light.
After this, the Battery LED (blue) lights. Please check
the ashing or lighting of each color.