Safety precautions
Table1(supportforIEC60601-1-2:2007andIEC60601-1-2:2014)
Guidelinesandmanufacture'sdeclaration-electromagneticemissions
This device is intended for use in the electromagnetic environment specied below.
The customer or the user of this device should assure that it is used in such an environment.
Emissions test Compliance Electromagneticenvironment-guidelines
RF emissions
CISPR 11
Group 1
The device uses RF energy only for its internal function. Therefore, its RF emissions
are very low and are not likely to cause any interference in nearby electronic equip-
ment.
RF emissions
CISPR 11
Class B
This device is suitable for use in all establishments including the following:
Domestic establishments and those directly connected to the public low-voltage
power supply network that supplies buildings for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage uctuations/icker
emissions
IEC 61000-3-3
Complies
IMPORTANT
• If I/F Cable4 1.5mD, I/F Cable4 8mD, and/or I/F Cable4 8mU is used, the emissions characteristics of this device is CISPR 11
class A that is suitable for the industrial areas and the hospitals. If it is used in a residential environment, for which CISPR 11
class B is normally required, this equipment might not oer adequate protection to radio-frequency communication services, and
it may be necessary to take a mitigation measures, for example, by relocating or re-orientating the device.
Table2(supportforIEC60601-1-2:2007)
Guidelinesandmanufacturer'sdeclaration-electromagneticimmunity
This device is intended for use in the electromagnetic environment specied below.
The customer or the user of this device should assure that it is used in such an environment.
Immunity test IEC60601testlevel Compliance level
Electromagneticenvironment-guide-
lines
Electrostatic discharge
(ESD)
IEC 61000-4-2
±6 kV contact ±6 kV contact
Floors should be wood, concrete or ceramic
tile. If oors are covered with synthetic
material, the relative humidity should be
at least 30 %. Mains power quality should
be that of a typical commercial or hospital
environment.
±8 kV air ±8 kV air
Electrical fast transient/
burst
IEC 61000-4-4
±2 kV for power supply
lines
±2 kV for power supply
lines
Mains power quality should be that of a typi-
cal commercial or hospital environment.
±1 kV for input/output lines ±1 kV for input/output lines
Surge
IEC 61000-4-5
±1 kV line(s) to line(s) ±1 kV line(s) to line(s)
±2 kV line(s) to earth ±2 kV line(s) to earth
Voltage dips, short
interruptions and voltage
variations on power sup-
ply input lines
IEC 61000-4-11
<5 % U
T
(>95 % dip in U
T
)
for 0.5 cycle
<5 % U
T
(>95 % dip in U
T
)
for 0.5 cycle
Mains power quality should be that of a
typical commercial or hospital environment.
If the user of the device requires continued
operation during power mains interruptions,
it is recommended that the device be pow-
ered from an uninterrupted power supply or
a battery.
40 % U
T
(60 % dip in U
T
)
for 5 cycles
40 % U
T
(60 % dip in U
T
)
for 5 cycles
70 % U
T
(30 % dip in U
T
)
for 25 cycles
70 % U
T
(30 % dip in U
T
)
for 25 cycles
<5 % U
T
(<95 % dip in U
T
)
for 5 sec
<5 % U
T
(<95 % dip in U
T
)
for 5 sec
Power frequency
(50/60 Hz) magnetic eld
IEC 61000-4-8
3 A/m 3 A/m
Power frequency magnetic elds should be
at levels characteristic of a typical location
in a typical commercial or hospital environ-
ment.
[NOTE] U
T
is the AC mains voltage prior to application of the test level.