Safety precautions
Table1(supportforIEC60601-1-2:2007andIEC60601-1-2:2014)
Guidelinesandmanufacture'sdeclaration-electromagneticemissions
This device is intended for use in the electromagnetic environment specied below.
The customer or the user of this device should assure that it is used in such an environment.
Emissions test Compliance Electromagneticenvironment-guidelines
RF emissions
CISPR 11
Group 1
The device uses RF energy only for its internal function. Therefore, its RF emissions
are very low and are not likely to cause any interference in nearby electronic equip-
ment.
RF emissions
CISPR 11
Class B
This device is suitable for use in all establishments including the following:
Domestic establishments and those directly connected to the public low-voltage
power supply network that supplies buildings for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage uctuations/icker
emissions
IEC 61000-3-3
Complies
IMPORTANT
If I/F Cable4 1.5mD, I/F Cable4 8mD, and/or I/F Cable4 8mU is used, the emissions characteristics of this device is CISPR 11
class A that is suitable for the industrial areas and the hospitals. If it is used in a residential environment, for which CISPR 11
class B is normally required, this equipment might not o󰀨er adequate protection to radio-frequency communication services, and
it may be necessary to take a mitigation measures, for example, by relocating or re-orientating the device.
Table2(supportforIEC60601-1-2:2007)
Guidelinesandmanufacturer'sdeclaration-electromagneticimmunity
This device is intended for use in the electromagnetic environment specied below.
The customer or the user of this device should assure that it is used in such an environment.
Immunity test IEC60601testlevel Compliance level
Electromagneticenvironment-guide-
lines
Electrostatic discharge
(ESD)
IEC 61000-4-2
±6 kV contact ±6 kV contact
Floors should be wood, concrete or ceramic
tile. If oors are covered with synthetic
material, the relative humidity should be
at least 30 %. Mains power quality should
be that of a typical commercial or hospital
environment.
±8 kV air ±8 kV air
Electrical fast transient/
burst
IEC 61000-4-4
±2 kV for power supply
lines
±2 kV for power supply
lines
Mains power quality should be that of a typi-
cal commercial or hospital environment.
±1 kV for input/output lines ±1 kV for input/output lines
Surge
IEC 61000-4-5
±1 kV line(s) to line(s) ±1 kV line(s) to line(s)
±2 kV line(s) to earth ±2 kV line(s) to earth
Voltage dips, short
interruptions and voltage
variations on power sup-
ply input lines
IEC 61000-4-11
<5 % U
T
(>95 % dip in U
T
)
for 0.5 cycle
<5 % U
T
(>95 % dip in U
T
)
for 0.5 cycle
Mains power quality should be that of a
typical commercial or hospital environment.
If the user of the device requires continued
operation during power mains interruptions,
it is recommended that the device be pow-
ered from an uninterrupted power supply or
a battery.
40 % U
T
(60 % dip in U
T
)
for 5 cycles
40 % U
T
(60 % dip in U
T
)
for 5 cycles
70 % U
T
(30 % dip in U
T
)
for 25 cycles
70 % U
T
(30 % dip in U
T
)
for 25 cycles
<5 % U
T
(<95 % dip in U
T
)
for 5 sec
<5 % U
T
(<95 % dip in U
T
)
for 5 sec
Power frequency
(50/60 Hz) magnetic eld
IEC 61000-4-8
3 A/m 3 A/m
Power frequency magnetic elds should be
at levels characteristic of a typical location
in a typical commercial or hospital environ-
ment.
[NOTE] U
T
is the AC mains voltage prior to application of the test level.