Users Manual Part 1

66
Operation of DR Detector
3�2�3 Exposure using Aero
Storage
IMPORTANT
The following DR Detector models can be used for Aero
Storage exposure: AeroDR 3 Detector.
Starting Aero Storage Mode
IMPORTANT
When DR Detector is in the following conditions, Aero Stor-
age mode cannot start:
It is connected to the AeroDR Battery Charger, AeroDR
Battery Charger2, and a wired cable.
DR Detector battery level has dropped to less than 10 %
level.
Selection SW cannot be operated while Aero Storage mode
is active.
When 100 Aero Storage images are saved, the Aero Stor-
age mode starts, but it turns to the sleep state, and expo-
sure using Aero Storage cannot be performed.
1
Press and hold the mode LED for three
seconds�
2
Ensure that the mode LED (blue) blinks
slowly,theinformationLEDindicatingthe
numberofsavedimagesison,andthe
status LED (blue) is on�
Status LED
(blue)
Information LED (white)
Mode LED
(blue)
Exiting Aero Storage Mode
IMPORTANT
You can also exit Aero Storage mode by pressing and
holding the mode LED for three seconds.
1
Insert DR Detector into the AeroDR Bat-
tery Charger and AeroDR Battery Char-
ger2,orconnectittoawiredcable.
2
AeroStoragemodeexits,andDRDetec-
tor is connected to the image processing
controller
HINT
Aero Storage mode automatically exits in the following
conditions:
DR Detector battery level drops to less than 3 % level.
DR Detector detects an error.
Exposure
IMPORTANT
If you do not plan to expose immediately, exit Aero Storage
mode.
In Aero Storage, the maximum exposure time is set to 1.8
seconds and the resolution is set to the default (200μm).
Please contact Konica Minolta technical representatives for
more information about how to change the settings.
The DR Detector is precision equipment, and therefore im-
pact or vibration during radiography or image transfer may
a󰀨ect the image quality. Be careful when handling the DR
Detector during and just after radiography.
The number of saved images displayed in the information
LED is counted when the orange status LED blinks fast af-
ter the X-ray exposure.
The number of saved images displayed after an exposure
is the patient’s image number.
Images exposed using Aero Storage do not have patient in-
formation. It is recommended, therefore, to keep record so
that you can associate the patient exposed with the number
of saved images displayed on DR Detector.
When the number of saved images in Aero Storage reach-
es 100, it turns to the sleep state, and exposure using Aero
Storage cannot be performed.
While importing the images, do not disconnect DR Detector
from AeroDR Battery Charger, AeroDR Battery Charger2,
or the wired cable.
When the image import process finishes, the number of
saved image in DR Detector becomes zero.
1
Start Aero Storage mode�
2
Check that this device is ready to expose
images,andthenpreparefortheexpo-
sure�