Various problems and countermeasures
5�2�5 Detector Interface Unit 2
Status Error description Corrective actions
LED(green/blue) does not light. AC power not supplied.
Make sure that the Ethernet cable is connect-
ed to the Power Supply Unit properly.
LED (blue) does not light.
The I/F cable is not connected to the DR De-
tector.
Make sure that the DR Detector is connected
to the I/F cable correctly.
The DR Detector is being used
with wired connection, but it is
not recognized on the image
processing controller.
The I/F cable is not connected to the DR De-
tector. Or, the Ethernet cable is not connected
to the Power Supply Unit.
Make sure that the DR Detector is connected
to the I/F cable correctly. Or, make sure that
the Ethernet cable is connected correctly.
Serial exposure does not start.
The Ethernet cable for relaying serial expo-
sure signals is not connected to the Sync port.
Make sure that the Ethernet cable for relying
serial exposure is connected to the Sync port.
5�2�6 Power Supply Unit
Status Error description Corrective actions
LED does not light. AC power not supplied.
Make sure the power cable is connected cor-
rectly.
The power switch cannot be
turned on.
There is an error in the power system.
Contact Konica Minolta technical representa-
tives.
The DR Detector is being used
with wired connection, but it is
not recognized on the image
processing controller.
The Ethernet cable is disconnected.
Make sure that the Ethernet cable is connect-
ed correctly.
5�2�7 AeroDR Generator Interface Unit
Status Error description Corrective actions
LED (green) does not light. AC power not supplied.
Make sure that the Ethernet cable is connect-
ed correctly. Or, make sure the power switch
of the AeroDR Interface Unit is on.
LED (green) does not change
from ashing to a solid light.
There is a communications error.
Contact Konica Minolta technical representa-
tives.