Charging of DR Detector
77
3�3�2 Charging with AeroDR
Battery Charger2
IMPORTANT
• Handle the DR Detector with extreme care when inserting it
into the AeroDR Battery Charger2.
• The wired connection connector of the DR Detector may
become warm right after charging on the AeroDR Battery
Charger2. This often occurs during charging and is not a
malfunction.
• If there is any problem during charging, the LED (orange)
on the AeroDR Battery Charger2 will light. Also, charging
will stop when an error occurs.
• As the DR Detector is locked by a magnet, do not pull it out
horizontally or vertically.
1
ConrmthattheLED(green)oftheAero-
DRBatteryCharger2isslowlyashing.
2
Slowly insert the DR Detector with its ex-
posure side pointed toward the operator
untilthebuzzersounds.Wheninserted,
charging of the DR Detector begins�
Exposure
side
Load slowly
Exposure
side
Inserted
3
Once the DR Detector is inserted correctly
andchargingstarts,theLED(blue)on
theAeroDRBatteryCharger2willashor
light�
• The LED (blue) on the AeroDR Battery Char-
ger2 changes according to the level of battery
power of the DR Detector.
4
Once charging of the DR Detector is com-
pleted,alltheLEDs(blue)ontheAeroDR
Battery Charger2 will turn on�
5
Remove the DR Detector from the AeroDR
Battery Charger2�
• Tilt the DR Detector in the direction shown in
the following gure and remove it.