Users Manual Part 2

1
2
3
4
5
6
7
8
9
10
11
12
133
Specications
7�1�18 SKR 3000
Item Description
Recommend-
ed storage
and usage
environment
conditions
When oper-
ating
Temperature Humidity Atmospheric pressure
10 °C to 35 °C
35 °C
10 °C
35 %RH to 85 %RH
(ensure no water
condensation)
35 %RH
85 %RH
700 hPa to 1060 hPa
* The temperature and humidity when operating the optional components, except Detector Interface
Unit and Power Supply Unit, are: 10 °C to 30 °C and 35 % to 80 %, respectively.
* For serial exposure, temperature: 10 °C to 30 °C, humidity 35 % to 85 %.
* The temperature and humidity when operating the Detector Interface Unit 2, Generator Interface
Unit 3, I/F Cable4 1.5mD, I/F Cable4 8mD and I/F Cable4 8mU, are: 10 °C to 30 °C and 35 % to
85 %, respectively.
* Limit continuous use in a hot and humid environment (35 °C to 37 °C/95 % or lower) of an incuba-
tor to within 25 minutes.
When not
operating
Temperature Humidity Atmospheric pressure
–10 °C to 40 °C
40 °C
–10 °C
20 %RH to 90 %RH
(ensure no water
condensation)
20 %RH
90 %RH
700 hPa to 1060 hPa
In storage/
transport
Temperature Humidity Atmospheric pressure
–20 °C to 50 °C
*1
-
20 °C
50 °C
20 %RH to 90 %RH
(ensure no water
condensation)
20 %RH
90 %RH
700 hPa to 1060 hPa
*1 However, performance warranty period when storing at 50 °C is 6 months after packing.
Classication Safety IEC60601-1 Class I
Operation
mode
Continuous operation