Switch User's Manual

-24-
7-2. The specification of control system connectors
7-2-1.Contact output connectors for power control (POWER CONT) and option (OPTION A-F)
button output
        relay (no polarity) relay (no polarity)
Relay type : photo MOS relay, make contact
Rating : maximum 50mA, less than AC/DC 24V
   Internal circuit is as follows.
 ◆Caution
Do not supply more than AC/DC 24V.
   When option button is set as momentary output, output pulse width is about 200msec.
fuse
varistor
A
B
C
D
E
F
OPTION OUT
POWER
CONT
Relay specification