Owner manual

Part C Technical data, Block diagram and Measuring principle Section 6
02/2001
ALTOFLUX 2W
41
%ORFNGLDJUDPRIVLJQDOFRQYHUWHU
1 AD converter
overdrive-proof signal processing,
processes flow peaks
up to 20 m/s (65ft/s) more rapidly and
accurately
than competition modules
digital signal processing and sequence
control
high-resolution Delta-Sigma A/D
converter,
digitally controlled and monitored
high signal-to-noise ratio, signal
processing: patent pending
2 Field power supply
The low-loss field power supply generates
the pulsed, electronically controlled DC
current for the magnetic coils of the
primary head
The field current ensures matched to the
power, an optimum signal amplitude and
signal-to-noise separation
3 Current output, HART
®
modem
galvanically isolated from all other
terminals, but not from power booster
converts the digital output signal from the
µP 2 microprocessor into a proportional
current
communication via HART®
4 Binary output
galvanically isolated from other groups
can be used as pulse or status output
output (B1), terminal for up to 100 mA
output (B2), terminal according NAMUR
(DIN 19 234)
5 Display/operator control unit
large LCD display, 3-line
3 keys for operator control of the signal
converter
connection to the internal IMoCom bus
6 IMoCom bus plug connector
for connection of external control and test
devices
7
Power booster input