ALTOSONIC V12 Handbook 12-chord ultrasonic gas flowmeter for custody transfer © KROHNE 04/2013 - 4002643502 - MA ALTOSONIC V12 R02 en
: IMPRINT ::::::::::::::::::::::::::::::::::::::: All rights reserved. It is prohibited to reproduce this documentation, or any part thereof, without the prior written authorisation of KROHNE Messtechnik GmbH. Subject to change without notice. Copyright 2013 by KROHNE Messtechnik GmbH - Ludwig-Krohne-Str. 5 - 47058 Duisburg (Germany) 2 www.krohne.
CONTENTS ALTOSONIC V12 1 Safety instructions 7 1.1 Intended use ..................................................................................................................... 7 1.2 Certifications .................................................................................................................... 7 1.3 Safety instructions from the manufacturer ..................................................................... 8 1.3.1 1.3.2 1.3.3 1.3.4 1.3.5 Copyright and data protection .......
CONTENTS ALTOSONIC V12 5 Start-up 31 5.1 Starting the signal converter ......................................................................................... 31 6 Operation 32 6.1 Display and operating elements .................................................................................... 32 6.1.1 Display in measuring mode with 2 or 3 measured values ................................................... 33 6.2 Menu overview .....................................................................
CONTENTS ALTOSONIC V12 8.9 Returning the device to the manufacturer..................................................................... 96 8.9.1 General information.............................................................................................................. 96 8.9.2 Form (for copying) to accompany a returned device............................................................ 97 8.10 Disposal .................................................................................................
CONTENTS ALTOSONIC V12 10.8.4 Holding Registers (read/write): Long integer (32-bit), addres range 5500-5999 ............ 132 10.8.5 Input Registers (read-only): Double (64-bit floating point), address range 6000-6499 .. 133 10.8.6 Holding Registers (read/write): Double (64-bit floating-point), addr. range 6500-6999. 133 10.8.7 Input Registers (read-only): Float (32-bit floating-point), address range 7000-7499 ..... 133 10.8.
SAFETY INSTRUCTIONS 1 ALTOSONIC V12 1.1 Intended use The ALTOSONIC V12 is a gas flowmeter for custody transfer applications. The meter is suitable to operate at least under the following conditions: • relative density from 0.55 and upwards • methane concentrations 75...100% CAUTION! High levels of CO2 can inhibit the operation of an ultrasonic flowmeter due to its acoustic absorption properties.
1 SAFETY INSTRUCTIONS ALTOSONIC V12 Other standards: • IECEx PTB 10.0013X INFORMATION! Not all country specific approvals are listed here. Please consult KROHNE in case of a specific approval that is not listed here. 1.3 Safety instructions from the manufacturer 1.3.1 Copyright and data protection The contents of this document have been created with great care. Nevertheless, we provide no guarantee that the contents are correct, complete or up-to-date.
SAFETY INSTRUCTIONS 1 ALTOSONIC V12 1.3.3 Product liability and warranty The operator shall bear responsibility for the suitability of the device for the specific purpose. The manufacturer accepts no liability for the consequences of misuse by the operator. Improper installation and operation of the devices (systems) will cause the warranty to be void. The respective "Standard Terms and Conditions" which form the basis for the sales contract shall also apply. 1.3.
1 SAFETY INSTRUCTIONS ALTOSONIC V12 1.3.5 Warnings and symbols used Safety warnings are indicated by the following symbols. DANGER! This information refers to the immediate danger when working with electricity. DANGER! This warning refers to the immediate danger of burns caused by heat or hot surfaces. DANGER! This warning refers to the immediate danger when using this device in a hazardous atmosphere. DANGER! These warnings must be observed without fail.
DEVICE DESCRIPTION 2 ALTOSONIC V12 2.1 Scope of delivery INFORMATION! Do a check of the packing list to make sure that you have all the elements given in the order. INFORMATION! The instrument is delivered in a seaworthy, reinforced wooden crate. Inspect the packing carefully for damages or signs of rough handling. Report damage to the carrier and to the local office of the manufacturer.
2 DEVICE DESCRIPTION ALTOSONIC V12 A series of acoustic transducer are installed on the inside of the meter body. Each pair of ultrasonic transducers forms an acoustic measurement path. The acoustic measurement path consists of one (direct) chord or two (reflective) chords. The chords in the horizontal plane are used for flow measurement, the chords in the vertical plane are used for diagnostics only.
DEVICE DESCRIPTION 2 ALTOSONIC V12 DANGER! The transducers have an Ex-d socket (socket, 3) that connects to an Ex-d plug 4 which terminates a coaxial wire 8. An alignment pin 5 on the connector engages with a slot in the Ex-d socket 2, this ensures that the transducer is connected with the correct polarity. The cap 6, screwed on the Ex-d socket, fastens the connector, a small securing screw (M2) 7 secures this cap.
2 DEVICE DESCRIPTION ALTOSONIC V12 Electronics compartment The compartment on the front has a cover with a glass window. It contains a frame with a number of printed circuit boards. The boards have the following functions: • • • • Frequency and status output board Processor boards Sensor driver board Display module (connected tot the front side of the frame, can be seen through the cover with the glass window) 2.
DEVICE DESCRIPTION 2 ALTOSONIC V12 Normally the meter is delivered with a set of parameters suitable for the application. In case a modification of the configuration file is made, this is stored in its data logging memory. This information can be retrieved afterwards for auditing and verification purposes. 2.4 Nameplates INFORMATION! Look at the device nameplate to ensure that the device is delivered according to your order. Check for the correct supply voltage printed on the nameplate.
2 DEVICE DESCRIPTION ALTOSONIC V12 Figure 2-7: Example of a FM nameplate Figure 2-8: Example of a CSA nameplate 16 www.krohne.
INSTALLATION 3 ALTOSONIC V12 3.1 Notes on installation CAUTION! Inspect the packaging carefully for damages or signs of rough handling. Report damage to the carrier and to the local office of the manufacturer. INFORMATION! Do a check of the packing list to make sure that you have all the elements given in the order. INFORMATION! Look at the device nameplate to ensure that the device is delivered according to your order. Check for the correct supply voltage printed on the nameplate. 3.
3 INSTALLATION ALTOSONIC V12 However for storage and/or transportation additional protective measures are recommended. Depending on how long the protection should last one of several methods can be considered: 1. For a storage period of less than one year, the inner pipewall should be protected by a corrosion inhibitor such as Shell Ensys. Do not apply the corrosion inhibitor on the acoustic transducers. 2. For a period of a year or longer, the pipewall should be protected by Tectyl or a similar product.
INSTALLATION 3 ALTOSONIC V12 3.4 Pre-installation requirements INFORMATION! The equipment is designed for safe operation under conditions according to the following classifications: • Pollution degree 2: this means that normally only nonconductive (dry) pollution will occur. Temporary conductivity caused by condensation can occur. • Protection class I: this means the equipment must be earthed. • Humidity: <95% RH • Ambient temperature: -40...+65°C / -40...+149°F • Suitable for indoor and outdoor use.
3 INSTALLATION ALTOSONIC V12 In order to install the gas flowmeter, the pipeline must have a slot of such length that the meter including the gaskets fits nicely in the slot. It should not be necessary to use excessive force to tighten the bolts in order to close the gaps on either side of the meter. Nor should the slot be too small, implying the slot has to be widened by applying brute force to fit the meter and gaskets in the slot.
INSTALLATION 3 ALTOSONIC V12 Without flow conditioner (AGA9, ISO 17089 and OIML R137 class 1) Figure 3-3: Reguired straight lengths for inlet and outlet 1 Inlet section: 5 DN 2 Outlet section: 3 DN With flow conditioner Figure 3-4: Reguired straight lengths for inlet and outlet 1 2 3 4 Inlet section before flow conditioner: 2 DN Flow conditioner (perforated plate) Inlet section after flow conditioner: 3 DN Outlet section: 3 DN INFORMATION! Contact KROHNE for recommendations on bi-directional use. 3.
3 INSTALLATION ALTOSONIC V12 3.5.6 P and T sensors Figure 3-5: Location of pressure and temperature transmitters 1 2 3 4 Install pressure transmitters on body of flowmeter at Pr point Install temperature transmitter at 2...5 DN downstream of flowmeter Install temperature transmitter under 45 degrees Install temperature transmitter with an insertion depth between 0.1 and 0.33 of nominal pipe diameter. INFORMATION! • See ISO 17089 for further details.
INSTALLATION 3 ALTOSONIC V12 3.6 Temperatures Figure 3-6: Temperatures 1 Ambient temperature 2 Process gas temperature 3 Use a sun shade to protect the flowmeter against direct solar radiation. CAUTION! SUNSHADE Direct solar radiation introduces temperature gradients in the metering section and must be avoided as much as possible. Use a sunshade or canopy over the flow, pressure and temperature transmitters for protection against direct exposure to sunshine.
4 ELECTRICAL CONNECTIONS ALTOSONIC V12 4.1 Safety instructions DANGER! All work on the electrical connections may only be carried out with the power disconnected. Take note of the voltage data on the nameplate! DANGER! Observe the national regulations for electrical installations! WARNING! Observe without fail the local occupational health and safety regulations. Any work done on the electrical components of the measuring device may only be carried out by properly trained specialists.
ALTOSONIC V12 ELECTRICAL CONNECTIONS 4 4.3 Digital I/O connections WARNING! 1. In order to prevent unauthorized or inadvertent opening and removal of the covers, an interlocking device is provided for each cover. Before a cover can be rotated (counter clockwise) for opening, release this interlocking device with a 2.5 mm Allen key. 2. The foot of the converter housing provide an earthing point, this must be connected to the nearest safety earth conductor. 3.
4 ELECTRICAL CONNECTIONS ALTOSONIC V12 Figure 4-3: Terminal compartment for inputs and outputs in field housing 1 Open the housing cover. 2 Push the prepared cable through the cable entry and connect the necessary conductors. 3 Connect the shield if necessary. • Close the cover of the terminal compartment. • Close the housing cover. INFORMATION! Each time a housing cover is opened, the thread should be cleaned and greased. Use only resinfree and acid-free grease.
ELECTRICAL CONNECTIONS 4 ALTOSONIC V12 4.3.3 Emulation of a turbine meter If the ultrasonic flowmeter should emulate a turbine meter, the following set up and settings can be implemented: • A/A-: Frequency output related to the line flow • B/B-: Frequency output inverted related to the line flow whereby this frequency output will stop operating if data valid alarm on status bit C/C- will occur. Place the frequency output B/B in series with status bit C/C- as presented in the figure shown below.
4 ELECTRICAL CONNECTIONS ALTOSONIC V12 4.4 Serial data communication (RS 485) Figure 4-5: Connection of serial data communication INFORMATION! For more information about Modbus, refer to Modbus protocol description and set-up on page 111. 4.5 Power connection WARNING! • Use a 24 VDC power supply to power the flowmeter, which complies to NEC class 2 (max. 100 VA, 24 VDC ±10% , see also IEC 61010-1, clause 6.3.1 and 6.3.2). The maximum power consumption is 17 W.
ELECTRICAL CONNECTIONS 4 ALTOSONIC V12 Figure 4-6: Location of power connector 4.6 Cabling WARNING! • Replace any unused cable gland by an Ex-d blind plug! • The temperature rating of all cables must have a temperature rating of at least 65°C. In case the process design temperature exceeds 65°C, the cables must have a temperature rating at least as high as the maximum process design temperature. WARNING! In respect of the model of the cable glands used, the outer diameter of the cable must be between 6.
4 ELECTRICAL CONNECTIONS ALTOSONIC V12 4.7 Grounding There are two screw connection points (one M5 thread and one M4 thread) to attach a ground conductor. They can be used to connect the upstream and downstream piping to the flowmeter (Equipotential). Figure 4-7: Location of grounding connectors 30 www.krohne.
START-UP 5 ALTOSONIC V12 5.1 Starting the signal converter The measuring device, consisting of the measuring sensor and the signal converter, is supplied ready for operation. All operating data have been set at the factory in accordance with your order specifications. CAUTION! The used power supply must be able to supply at least 3 A during the start-up of the converter. When the power is switched on, a self test is carried out (KROHNE logo is visible during the test).
6 OPERATION ALTOSONIC V12 6.
OPERATION 6 ALTOSONIC V12 Key Measuring mode Menu mode Submenu or function mode Parameter and data mode > Switch from measuring mode to menu mode; press key for 2.
6 OPERATION ALTOSONIC V12 6.2 Menu overview Menu number Description 4.1.2 Settings 4.1.2.1 Language 4.1.2.2 Default Page 4.1.2.3 Bargraph Labrl 4.1.3 Measurement Page 1 4.1.3.1 Number of lines 4.1.3.2 Line Settings 1 4.1.3.3 Line Settings 2 4.1.3.4 Line Settings 3 4.1.4 Measurement Page 2 4.1.4.1 Number of lines 4.1.4.2 Line Settings 1 4.1.4.3 Line Settings 2 4.1.4.4 Line Settings 3 4.1.5 Graphical Page 4.1.5.1 Show Value 4.1.5.2 Graph type 4.1.5.3 Minimum 4.1.5.
OPERATION 6 ALTOSONIC V12 6.3 Access control and seals All acoustic transducers and the electronics unit of ALTOSONIC V12 can be exchanged without the need for recalibration. The only exception is the board that holds all Custody Transfer (CT) settings. After initial calibration of the ALTOSONIC V12 all CT settings are stored inside the meter and a dipswitch no 4 is put in the locked position.
6 OPERATION ALTOSONIC V12 INFORMATION! The sealing of the cover is not required and can be broken without affecting the CT approval. The nameplate must be sealed to the flowmeter, as shown in the figure below. Alternatively the nameplate can be welded on the flowmeter or be mounted with rivets, so that removal without destroying the name plate is not possible. Figure 6-5: Sealing of nameplate 36 www.krohne.
SOFTWARE SERVICE TOOL 7 ALTOSONIC V12 7.1 Introduction The KROHNE Flowmeter Configuration and Monitoring Tool is a software package to support the application of the ALTOSONIC V12 ultrasonic gas flowmeters. It is designed to be used with a PC with a Windows operating system and can be downloaded from www.krohne.com. It can: • collect data from a flowmeter • present data from a flowmeter • verify / set / adjust parameters used by the software inside the flowmeter.
7 SOFTWARE SERVICE TOOL ALTOSONIC V12 Figure 7-2: Start-up screen, lower part The first field is an information field, reserved to show the communication method in operation. Without an active communication the first and the second fields will show a dash. When communication with a flowmeter has been established, the second field will show the address of this flowmeter, valid for the communication method in operation. Figure 7-3: Connect the device 1. Choose and click the menu pull down button Device.
SOFTWARE SERVICE TOOL 7 ALTOSONIC V12 KROHNE Flowmeter Configuration and Monitoring Tool will now send a message, asking for a response from any meter that is able to communicate using the selected communication method. Figure 7-5: Settings for Modbus port 0 04/2013 - 4002643502 - MA ALTOSONIC V12 R02 en www.krohne.
7 SOFTWARE SERVICE TOOL ALTOSONIC V12 Figure 7-6: Settings for Modbus port1 When using TCP/IP (network environment) a dialog box will appear listing the responding meters. If multiple instruments are present in the same network a list will be presented. The other communication option are peer-to-peer and will request for setting up the communication link. An arrow in the most left column in the dialog box marks the meter selected to start a communication session. 40 www.krohne.
SOFTWARE SERVICE TOOL 7 ALTOSONIC V12 Figure 7-7: Select the device 1. To confirm the selection, click on the arrow in the most left field of the row with the meter selected by default. 2. To select another meter: move the arrow to the row with the meter you want to communicatie with, click in the most left field of the row showing the meter. 3. Confirm your choice by clicking OK.
7 SOFTWARE SERVICE TOOL ALTOSONIC V12 Figure 7-9: Login dialog box Standard: User: operator / Password: operator or User: supervisor / Password: supervisor The selected communication method and flowmeter address will appear in the respective information fields in the status bar at the bottom of the screen. Figure 7-10: Status bar information 7.4 Loading a monitoring configuration INFORMATION! There are two different methods to obtain data from a flowmeter.
SOFTWARE SERVICE TOOL 7 ALTOSONIC V12 Step 1: Click the button “File” to open the File menu. Step 2: Select the option “Open Monitoring Configuration”. Figure 7-11: File menu A browser window will open. Figure 7-12: Open monitoring configuration Step 3: Select a monitoring configuration file and click “Open”. INFORMATION! You can create and save one or more monitoring configurations of your own, according to your own preferences. Later it will be explained how you can do this.
7 SOFTWARE SERVICE TOOL ALTOSONIC V12 The monitoring status field updates from “Mon.OFF” to “Mon.ON”, and turns from yellow into green. Figure 7-13: Green status field Alternatively you can configure the software to automatically find a monitoring configuration file and start monitoring immediately after the program is launched. Step 5: Click the “Tools” button to open the tools menu. Figure 7-14: Tools menu Step 6: Click “Settings”. A window “Settings” with 4 tabbed sheets will open.
SOFTWARE SERVICE TOOL 7 ALTOSONIC V12 Step 8: You can: • click the button “Use current configuration”, or • click “Browse” to open another window This is an identical browse window as used before to find a monitoring configuration file. Select the file you want to use to start monitoring the next time you connect to a meter. Step 9: Click “OK” to confirm the settings. The next time you launch the program the steps 1...4 will be skipped. 7.5 Changing and saving a monitoring configuration 7.5.
7 SOFTWARE SERVICE TOOL ALTOSONIC V12 Step 3: Click “Yes” to stop the monitoring activity. INFORMATION! Note that the instrument itself will not stop measuring! Only the presentation of measuring results will be interrupted. The Monitoring Configuration dialog appears: Figure 7-18: Monitoring configuration Step 4: Click “Edit List” . The Monitoring List editor dialog appears: Figure 7-19: Monitoring list 46 www.krohne.
SOFTWARE SERVICE TOOL 7 ALTOSONIC V12 Variables can be added to the list of values to be collected from the meter and become available for presentation: Step 5: Open the tree structure in the left panel, and select the variable name to be added to the list. Step 6: Click the button pointing to the right in the center panel. The item will now appear in the list in the right panel.
7 SOFTWARE SERVICE TOOL ALTOSONIC V12 Figure 7-21: Restart of monitoring Step 14: Click “Yes” to resume the monitoring process. 7.5.2 Saving a monitoring configuration with a new name INFORMATION! • If you want to save both the modified monitoring configuration as well as the monitoring configuration you used as a starting point, you will need to save now the modified configuration (with a new unique name).
SOFTWARE SERVICE TOOL 7 ALTOSONIC V12 Figure 7-23: Save monitoring configuration Step 3: Type a new name in the box “Filename”. Step 4: Click "Save". Your new monitoring configuration is now saved with its new name. 7.5.3 Saving a monitoring configuration with its current name You can save an updated version of your monitoring configuration at any time: Step 1: Click the "File" button to open the file menu. Figure 7-24: File menu Step 2: Click “Save Monitoring Configuration”.
7 SOFTWARE SERVICE TOOL ALTOSONIC V12 7.6 Creating a monitoring configuration CAUTION! If you have no monitoring configuration file available to start with, or when you have not yet opened a monitoring configuration file, the option “Configure…” in the “Monitoring” menu is not available. You have to use the option “New Monitoring Configuration…” in the File menu to create a monitoring configuration file. Step 1: Click the “File” menu button.
SOFTWARE SERVICE TOOL 7 ALTOSONIC V12 Figure 7-27: Monitoring configuration Step 3: Click “Edit List”. The empty Monitoring List Editor window will appear. Figure 7-28: Monitoring list editor Step 4: Open the tree in the panel to the left of the window, select the data items you want to monitor and add these to the list in the panel to the right of the window, by pressing the ">" button. Step 5: When finished, click “Close” in the “Monitoring List Editor” window.
7 SOFTWARE SERVICE TOOL ALTOSONIC V12 Step 7: Click “Close” in the “Monitoring Configuration” window. The “Monitoring” window reappears: Figure 7-29: Start monitoring? Step 8: Click “Yes” to start the monitoring process. Store the new monitoring configuration using the “Save…” option in the File menu as described before. Assign a name to the new monitoring configuration: the “Save Monitoring Configuration” screen will appear.
SOFTWARE SERVICE TOOL 7 ALTOSONIC V12 This window may contain multiple sheets or tabs. Each tab represents a sheet displaying a selection of data, in numerical or graphical format (or both). More sheets may be present allowing to present different sets of data or to present data in different formats. A particular sheet is identified by means of its name as is shown on its Tab. 3. In the window User Views, any sheet may be selected by clicking on its associated Tab. 4.
7 SOFTWARE SERVICE TOOL ALTOSONIC V12 7.8 Customizing the way data is presented The data presented in “User Views” can be configured according to your speciific needs or preferences. This applies to the selection of the data and to the way the data is presented: in numerical format or as a graph. INFORMATION! Only data that is actually gathered from the flowmeter (as set up in the Monitoring Configuration), is available to be displayed.
SOFTWARE SERVICE TOOL 7 ALTOSONIC V12 CAUTION! The files used for the preset “User Views” tabs can be reused. This is the easiest way to create your own “sheets” or “tabs” : making combinations of the predefined grids and/or graphs according to your preference. In case you use an existing – predefined- format (even if it appeared before as a format filling only a half sized upper or lower panel) to fill a sheet or tab with only one panel, it will automatically expand to fill the whole of the panel.
7 SOFTWARE SERVICE TOOL ALTOSONIC V12 Figure 7-33: Add tab Step 5: Click “Grid”, “Trend” or “X-Y Graph” according to your choice and then click “OK”. Step 6: Click the “down” arrow to the right of the box configuration file. Step 7: A list of available presentation formats will become visible. Click the file of your choice. Step 8: If desired, select a format and a definition file for the second panel. Step 9: Click the “OK” button. Now the new sheet will appear with one or two panels as you have chosen.
SOFTWARE SERVICE TOOL 7 ALTOSONIC V12 Figure 7-34: Add columns Step 6: Enter the number of columns in the dialogue box that appears and click OK. Figure 7-35: Add columns Step 7: Click again in the gray area with the right mouse button to open the context menu, and click the option “Add Row(s)”. Figure 7-36: Add Rows Step 8: Type the number of rows in the dialogue box that appears and click OK. 04/2013 - 4002643502 - MA ALTOSONIC V12 R02 en www.krohne.
7 SOFTWARE SERVICE TOOL ALTOSONIC V12 Figure 7-37: Add rows The size of the grid is now defined, however it can be adjusted at any time adding columns or rows. To define the information to be displayed by a specific cell: Step 1: Position the cursor on the cell to be defined. INFORMATION! Do not forget the next step! If you skip the cell selection and continue with the following steps you might edit the content or format a cell you don't want to change.
SOFTWARE SERVICE TOOL 7 ALTOSONIC V12 Step 11: For the box referenced as “Property”, open the list to show the properties that can be selected: Label / Fieldname / Name / Full path / Parent name / Full parent path / Actual value / Average value / Minimum value / Maximum value / Unit / Status. Step 12: Click the property to be displayed. Step 13: Click ”OK”. Alternative way for steps 6...8: Step 6: Click “User text”. Step 7: Type the text string to be displayed in the cell in the field “User text”.
7 SOFTWARE SERVICE TOOL ALTOSONIC V12 7.8.3 Creating a new graphical presentation Step 1: Click “Add Tab”. Step 2: Type a name for the new tab. Step 3: In the box “Type” click on the option “Trend”. Note: Leave the “Configuration File” field empty. Step 4: Click “OK”. An empty graphic screen appears. Step 5: Click the button “Configure”. A window will open for defining the settings of the graphical presentation.
SOFTWARE SERVICE TOOL 7 ALTOSONIC V12 7.9 Creating reports The service tool has been developed to also enable you to create records regarding the state of the ultrasonic flowmeter and the way the meter performs. Records can be printed reports or records can be exported and saved as data files on a storage medium.
7 SOFTWARE SERVICE TOOL ALTOSONIC V12 Figure 7-39: Parameters Use the scroll bar at the right side of the window to scroll through the list and the scroll bar at the bottom to display the columns you are interested in. Creating a formatted report Go to "Reports & Exports > Reports > Standard parameter report". Figure 7-40: Parameter report A formatted preview of the Standard Parameter Report will be displayed: 62 www.krohne.
SOFTWARE SERVICE TOOL 7 ALTOSONIC V12 Figure 7-41: Preview of report As this report will occupy several pages you can use the buttons on top of the screen to select a particular page. The small panel at the left side lists – as a tree structure - all the items included in the report. You can use the scroll bar in the right of the small panel to position immediately the report page where this parameter is printed in front in the right panel.
7 SOFTWARE SERVICE TOOL ALTOSONIC V12 Exporting a formatted report Open the Report Preview’s Export menu: Figure 7-43: Report preview Select the application format of the exported report: Adobe Acrobat (.pdf), Microsoft Excel (.xls), or Microsoft Word (.doc). use the file selector dialogue to select the location and the filename for the export. Create a file listing the parameters in CSV format Click “Tools” to open the Tools menu. Click “Reporting” to open the Reporting submenu.
SOFTWARE SERVICE TOOL 7 ALTOSONIC V12 Figure 7-45: Export parameter overview Click the button “Export”. In the output dialog box that appears: Specify the location where you want to store the file. Type a unique file name for the document. Click the button “Save”. Saving a parameter file in .XML format Saving and storing the parameter file as in .XML format is an important feature.
7 SOFTWARE SERVICE TOOL ALTOSONIC V12 7.9.2 Reporting related to process values For the reporting functions related to actual process values the same features and procedures are applicable as valid for the parameters. 7.9.3 Reporting related to calibration parameters The calibration parameters is a subset of the whole collection of parameters. The calibration parameters are those parameters that will effect the value of the measuring results.
SOFTWARE SERVICE TOOL 7 ALTOSONIC V12 Click the option “Logging”. The check mark will now disappear, the field in the status bar at the bottom of the screen will become yellow and display “Log.OFF”: Figure 7-47: Yellow color of Log.OFF To restart the logging process: Open the “Monitoring” menu. Click “Logging”. The check mark will reappear and the status field will display “Log.ON” and turn to green again. Logged data will be stored in a file for each day (provided it is the same meter).
7 SOFTWARE SERVICE TOOL ALTOSONIC V12 Figure 7-48: Monitoring configuration Click the down arrow to the right of the box log rate. Select a value. Example (as displayed above): When the “Fetch Rate” equals 2, the set of data collected from the flowmeter is refreshed every 2 seconds. When the “Log Rate” is set to 5, the fifth of each set of data collected from the flowmeter will be stored in the log file. Effectively this means a set of data is stored in the log file every 10 seconds.
SOFTWARE SERVICE TOOL 7 ALTOSONIC V12 The window “Monitoring List Editor” will open: Figure 7-49: Monitoring List Editor The variables listed in the window are the variables selected for being collected for monitoring purposes. INFORMATION! • It is possible to store different representations of any variable: the actual value, the average value, the minimum value or the maximum value.
7 SOFTWARE SERVICE TOOL ALTOSONIC V12 Step 5: Click on the variable to select it. Step 6: Click on the button pointing to the right in the panel in the middle. Step 7: Put check marks in the columns for the representations you want to be logged. Step 8: Click the Monitoring List Editor’s “Close” button. Step 9: Click the Monitoring Configuration dialogue’s “Close” button to accept the configuration.
SOFTWARE SERVICE TOOL 7 ALTOSONIC V12 For Duration, select a time span of: • • • • • 1 hour 4 hours 12 hours 1 day 1 week Click on your choice in the value list associated with the box “ Timespan”. New intervals start at multiples of the selected “Timespan”. In case of week the start of a new interval (and the end of the previous interval) occurs on the day specified by “Week starts at”. When using “Number of records”, specify any number of records in the box “Max.Records”.
7 SOFTWARE SERVICE TOOL ALTOSONIC V12 Figure 7-51: Settings Step 3: 3:Type in the box “Logging Home Folder” the name (and path) of the directory where you want your data to be stored. 7.12 Adjusting parameter settings A typical adjustment of the parameter setting is the adjustment of the meter factor. The meter factor is determined during the calibration and is programmed before the ultrasonic flowmeter is sealed.
SERVICE 8 ALTOSONIC V12 8.1 Periodic maintenance The ultrasonic gas flowmeter does not contain any moving parts. During operation no maintenance will be required such as lubrication or replacing parts that wear out. Therefore during normal operation there will be no need to disassemble the meter in order to gain access to the inner parts of the meter. It is however good practice to make a log file on regular basis, refer to Logging data from a flowmeter on page 66.
8 SERVICE ALTOSONIC V12 8.3 Exchange of transducers Transducers are not likely to break down. In the unlikely situation that a transducer fails, the transducer has to be replaced. Although not strictly required, it is common practice to replace the other transducer on the same acoustic path at the same time. If possible, depressurise the flowmeter. If this is not possible, use the extraction tool to remove the transducer under pressure (maximum 150 bar(g)).
SERVICE 8 ALTOSONIC V12 Figure 8-2: Transducer types 1 Epoxy type 2 Titanium type Figure 8-3: Delrin protection cap 1 Delrin protection cap (only at welded meter bodies) Steps to remove a transducer: • Remove the Delrin protection cap (only for welded meter bodies). • Untighten and release the small M2 screw (number 1 in picture below). Figure 8-4: View at installed transducer 1 Hexagonal securing screw M2 2 Holes for removal of cap 04/2013 - 4002643502 - MA ALTOSONIC V12 R02 en www.krohne.
8 SERVICE ALTOSONIC V12 Figure 8-5: Remove cap • Install the special tool 1 with the two pins 2 in the cap 3. • Turn the special tool counter clockwise 4 to loosen the cap from the transducer. Make sure that the cable is not twisted. • Pull the coax cable with the Ex-d plug out of the transducer.
SERVICE 8 ALTOSONIC V12 Figure 8-8: Pulling out the transducer 1 Pull to remove the transducer 2 Turn the transducer to loosen the transducer from the tool Steps to install a transducer: CAUTION! • Make sure that the inside of the transducer nozzle is completely clean. • Make sure that the two O-rings of the new transducer are a little greasy. Figure 8-9: Installing transducer in tool 1 Put the tool on the transducer. 2 Turn the tool clockwise to attach it to the transducer.
8 SERVICE ALTOSONIC V12 • Plug the coax cable connector back into the transducer.
SERVICE 8 ALTOSONIC V12 8.5 Exchange of transducers - Pressurised condition WARNING! The tool is designed to be able to replace a transducer of a ultrasonic gas flowmeter - while it is pressurized - in a safe way. Nevertheless, working with equipment with high gas pressure requires utmost care.
8 SERVICE ALTOSONIC V12 INFORMATION! • Whenever it is possible to isolate and depressurize an ultrasonic gas flowmeter in case it would be necessary to replace one or more transducers, that should be the preferred way to proceed instead of using the retraction tool. • Decide which persons need to be involved in the operation, especially the ones that have to perform the job at the location of the meter, empower procedures to limit access to the area where the operation takes place only to them. 80 www.
SERVICE 8 ALTOSONIC V12 8.5.1 Transducer retraction tool Figure 8-14: Retraction tool 04/2013 - 4002643502 - MA ALTOSONIC V12 R02 en www.krohne.
8 SERVICE ALTOSONIC V12 Number in picture Description 1 Hand wheel at upper end of the main shaft 2 Adjusting disk to position the shaft 3 Locking device 4 Main shaft 5 Tube 6 Pressure relief valve 7 Clamping device 8 Handle to operate the ball valve 9 Ball valve 10 Intermediate flange Only to be used for flowmeters with welded transducers 11 Wedge shape adapter Only to be used for flowmeters machined out of one piece 12 Spring loaded hexagon at lower end of main shaft 13 Nut to
SERVICE 8 ALTOSONIC V12 Figure 8-15: Transducers on front side of the flowmeter Four transducers (belonging to two acoustic paths) are located on the back side of the flowmeter. The codes and locations of these transducers are indicated in the figure below. Figure 8-16: Transducers on back side of the flowmeter Two transducers are located on the top of the ultrasonic flowmeter. The codes and locations of these transducers are indicated in the figure below.
8 SERVICE ALTOSONIC V12 Step 2: Disconnect the transducer cable INFORMATION! The transducer cable must be disconnected before the retraction tool can be used. Disconnect the transducer connection cable in the following steps. • Untighten and release the small M2 screw (number 7 in picture below) on top of the cap fastened to the extension with M10 screw thread on top of the transducer.
SERVICE 8 ALTOSONIC V12 Step 3: Test the suspected transducer INFORMATION! Before starting the operation to replace a transducer care must be taken to make sure which transducer is suspect to be failing. As a start the diagnostic tool to check the performance of the ultrasonic gas flowmeter shall be used for this purpose. In case of doubt consult the local support organisation of the manufacturer. Be aware transducers always operate as pairs (belonging to the same acoustic path).
8 SERVICE ALTOSONIC V12 Figure 8-21: Mount bronze fixation nut 1 2 3 4 Transducer and fastening nut inside the flowmeter Mounted transducer adapter. Bronze nut Use a 5 mm hexagon key. • Fasten the transducer adapter by means of the bronze nut using a 5mm hexagon key. Step 5: Mount the adapter plate INFORMATION! To be able to install the retraction tool on the transducer port, a transducer adapter plate must be used. Two different models are available.
SERVICE 8 ALTOSONIC V12 CAUTION! The square wedge adapter plate is available with two offset angles. For flowmeters with a nominal size of 6", the adapter must have an offset angle of 18 degrees. For flowmeters with a nominal size of 8" and larger, the adapter must have an offset angle of 20 degrees. • Fasten the square wedge shaped adapter by means of 4 bolts M8 x 25 (grade 12.9) with a 6 mm hexagon key. Mind the proper angle of the adapter. • Proceed with Step 6: Mount the retraction tool tool.
8 SERVICE ALTOSONIC V12 INFORMATION! When properly tightened there should still be some clearance between both plates of the adapter. Step 6: Mount the retraction tool • Fasten the lower part of the retraction tool (the ball valve) on the transducer adapter by means of 4 bolts M 8 x 130 (grade 12.9) using the 6 mm hexagon screw driver. Check the O-ring sealing, if damaged replace (size 40.87 x 3.53 mm). • Tighten the M8 bolts with appropriate force using the 6 mm hexagon angled key.
SERVICE 8 ALTOSONIC V12 • Measure the position of the end of the hexagon on the main shaft relative to the flange of the tube. • Memorize and note this value as "Length". • Make sure that the knob of the locking device is lowered to disable rotation of the adjusting disk. • Turn the handwheel to adjust this "Length" to a value of "Depth" minus 35 mm. Figure 8-26: Measure length 1 Put the ruler to the flange of the tube. 2 Read the length at the end of the hexagon.
8 SERVICE ALTOSONIC V12 Figure 8-28: Clamping device Step 7: Extracting the transducer • Turn the hand wheel a few (one or two) turns clockwise until resistance is ensed as a result of the hexagon at the end of the shaft becoming engaged with the adapter. • Use the ruler to measure the position of the shaft relative to the tube (the distance between the inner side of the hand wheel and the adjusting disk). • Memorize and note this value as "Reference".
SERVICE 8 ALTOSONIC V12 • Memorize and note the value : "Lowest shaft position" (when a transducer is placed back again this value will be needed to check that the transducer has arrived at the correct position. • Turn the knob of the locking device to allow it to return to its locked position. • Turn the adjusting disk back (clockwise) to find the position where the locking device will lock.
8 SERVICE ALTOSONIC V12 • Close the ball valve. • Open the pressure relief valve to allow the gas in the cavity between the ball valve and the tube to escape. Figure 8-32: Close ball valve 1 Close ball valve 2 Open pressure relief valve • Untighten and remove the bolts of the clamping device. • Remove both halves of the clamping device from the combination of the tube and the ball valve. INFORMATION! Due to friction this may require some force.
SERVICE 8 ALTOSONIC V12 • Turn the hand wheel clockwise to move the end of the main shaft with the transducer outward. The assembly of the transducer and the adapter is now accessible to be removed from the shaft. Figure 8-34: Extracted transducer 1 Turn wheel clockwise 2 Extracted transducer, fastening nut and transducer adapter The removed transducer can now be detached from the transducer adapter.
8 SERVICE ALTOSONIC V12 8.6 Exchange of electronics unit Before opening the converter housing: DANGER! All work on the electrical connections may only be carried out with the power disconnected. Take note of the voltage data on the nameplate! WARNING! Observe without fail the local occupational health and safety regulations. Any work done on the electrical components of the measuring device may only be carried out by properly trained specialists.
SERVICE 8 ALTOSONIC V12 • Loosen all coax cables on the bottom of the electronic unit (the number of cables depend on the version of the flowmeter). CAUTION! Please pay attention that the same amount of force is applied on both pullers, otherwise the connector at the backside can be damaged. Figure 8-36: Location of connectors 1 Position of connectors for transducer cables 2 Metal pullers DANGER! Electrostatic discharge (ESD) can damage electronic parts.
8 SERVICE ALTOSONIC V12 8.7 Battery maintenance Within the electronics a battery is present. This battery is used only for the internal clock. If a meter is installed and operating from an external power supply the battery will not be used. If a meter is shut down or on stock than the battery will power the internal clock. The following maintenance actions with respect to the battery should be taken: • Replace the battery after 10 years of service.
SERVICE 8 ALTOSONIC V12 8.9.2 Form (for copying) to accompany a returned device Company: Address: Department: Name: Tel. no.: Fax no.: Manufacturer's order no. or serial no.: The device has been operated with the following medium: This medium is: water-hazardous toxic caustic flammable We checked that all cavities in the device are free from such substances. We have flushed out and neutralized all cavities in the device.
9 TECHNICAL DATA ALTOSONIC V12 9.1 Measuring principle The ultrasonic gas flowmeter operates according to the principle of measuring the transit time of an ultrasonic sound wave. A gas velocity is derived from the difference in transit time of a sound wave travelling in a direction with the flow direction and the sound wave travelling in the opposite direction. The trajectory of the sound wave is called the acoustic path. A chord is the direct path crossing the pipe from one side to the opposite side.
TECHNICAL DATA 9 ALTOSONIC V12 The velocity of the gas is derived from formula (1) and (2) as: An important feature of this method is that the calculated gas velocity does not depend on the speed of sound in the gas or gas properties in general. The gas velocity as calculated is only a function of the measured transit times tAB and tBA; the length of the chord and the angle of intersection of the measuring chord are supposed to be known from the design of the flowmeter.
9 TECHNICAL DATA ALTOSONIC V12 Figure 9-3: Swirl pattern after a single bend Swirl can disturb the gas velocity measurement as the tangential velocity component adds up with the gas velocity measured along one chord. However chords can be arranged in a way to compensate for swirl effects. Two chords arranged in the same plane intersect the flow at opposite angles (cross configuration). These chords respond to the tangential velocity component with the same magnitude but with the opposite sign.
TECHNICAL DATA 9 ALTOSONIC V12 9.5 Technical data table INFORMATION! • The following data is provided for general applications. If you require data that is more relevant to your specific application, please contact us or your local sales office. • Additional information (certificates, special tools, software,...) and complete product documentation can be downloaded free of charge from the website (Download Center).
9 TECHNICAL DATA ALTOSONIC V12 Display and user interface Graphic display LC display, backlit white Size: 128x64 pixels, corresponds to 59x31 mm = 2.32"x1.22". Display turnable in 90° steps. The readability of the display could be reduced at ambient temperatures below -25°C / -13°F. Operator input elements 4 optical keys for operator control of the signal converter without opening the housing.
TECHNICAL DATA 9 ALTOSONIC V12 Pressure Pressure range 1...450 bar / 0.1...45MPa / 15...6525 psi (ASME 150...2500) All sensor designs at full rating acc. to below flange standards for standard materials. Properties of medium (other properties on request) Physical condition Natural gas with a minimum of 75% methane. other applications on request. Wet gas content Typically ≤ 1% LVF, contact manufacturer for detailed sizing.
9 TECHNICAL DATA ALTOSONIC V12 Inputs and outputs MODBUS Description Modbus RTU or Modbus ASCII, Slave, RS485 (galvanically isolated) Transmission procedure Half duplex, asynchronous Address range 1…247 Supported function codes 03, 04, 06, 08, 16 Supported Baudrate 50, 75. 110, 150, 300, 600, 1200, 2400, 4800, 9600, 19200, 38400, 56000, 64000, 115200, 128000 Baud Approvals and certificates CE This device fulfills the statutory requirements of the EC directives.
TECHNICAL DATA 9 ALTOSONIC V12 9.6 Dimensions and weights INFORMATION! • Flowmeters with diameters ≥ 6" and ASME ≤ 900 lb are standard equipped with transducers that are retractable under pressure. • All measures are provided as indication. They can vary slightly with different schedule sizes. • Values for larger diameters are avalable on request.
9 TECHNICAL DATA ALTOSONIC V12 ASME 300 lb Nominal size [mm] H [mm] L [mm] Weight [kg] 100 520 400 1 158 150 570 450 248 200 620 600 371 250 680 750 533 300 760 900 755 350 810 1050 1008 400 870 1200 520 450 920 1350 659 500 980 1500 862 600 1100 1800 1354 1 300 mm available on request. ASME 300 lb Nominal size [inch] H [inch] L [inch] Weight [lbs] 4 20.47 15.75 1 348 6 22.44 17.72 547 8 24.41 23.62 818 10 26.77 29.53 1175 12 29.92 35.
TECHNICAL DATA 9 ALTOSONIC V12 ASME 600 lb Nominal size [mm] H [mm] L [mm] Weight [kg] 100 520 400 1 168 150 575 450 271 200 630 600 411 250 710 750 618 300 780 900 850 350 815 1050 1070 400 880 1200 640 450 930 1350 805 500 1000 1500 1055 600 1100 1800 1621 1 300 mm available on request. ASME 600 lb Nominal size [inch] H [inch] L [inch] Weight [lbs] 4 20.47 15.75 1 370 6 22.64 17.72 598 8 24.8 23.62 906 10 27.95 29.53 1363 12 30.71 35.
9 TECHNICAL DATA ALTOSONIC V12 ASME 900 lb Nominal size [mm] H [mm] L [mm] Weight [kg] 100 1 520 400 176 150 590 600 324 200 660 600 464 250 730 750 684 300 810 900 957 350 840 1050 1190 400 890 1200 720 450 960 1350 964 500 1020 1500 1254 600 1160 1800 2200 1 Minimum Inner diameter: 80 mm (≤ sch 80). ASME 900 lb Nominal size [inch] H [inch] L [inch] Weight [lbs] 41 20.47 15.75 388 6 23.23 23.62 714 8 25.98 23.62 1023 10 28.74 29.
TECHNICAL DATA 9 ALTOSONIC V12 ASME 1500 lb Nominal size [mm] H [mm] L [mm] Weight [kg] 100 1 530 500 221 150 600 600 434 200 660 800 652 250 760 750 1030 300 860 900 1507 1 Minimum Inner diameter: 80 mm (≤ sch 80). ASME 1500 lb Nominal size [inch] H [inch] L [inch] Weight [lbs] 41 20.87 19.69 487 6 23.62 23.62 957 8 25.98 31.5 1438 10 29.92 29.53 2271 12 33.86 35.43 3323 1 Minimum Inner diameter: 3.15" (≤ sch 80).
9 TECHNICAL DATA ALTOSONIC V12 9.
ALTOSONIC V12 MODBUS PROTOCOL DESCRIPTION AND SET-UP 10 10.1 Introduction Modbus communication is based on the master-slave principle. Only the master can initiate transactions (requests), and only the addressed device (slave) responds. The master can also send a broadcast message (“message to all”); none of the slaves will respond to such a message. The flowmeter always acts as a Modbus compatible slave when communicating with host systems. Slaves are identified by means of a “device address”.
10 MODBUS PROTOCOL DESCRIPTION AND SET-UP ALTOSONIC V12 Multiple flowmeters may be connected to the same RS 485 line. In this case only the terminating resistor in the flowmeter at the end of the line should be connected. The terminating resistors in the other flowmeters on the line should be disconnected by means of the switch on the RS 485 driver printed circuit board. Default this switch is set to connect the line terminating resistor.
ALTOSONIC V12 MODBUS PROTOCOL DESCRIPTION AND SET-UP 10 10.3.2 RTU mode Each byte of data is represented in the message by an equivalent number of bits (8). The number of bits transmitted in the process of communicating one byte of information is sometimes also referred to as a “character”. Note that this is not the same as an ASCII character. Standard serial communication parameters Start bits 1 Data bits 8 Parity even Stop bits 1 Error check field Cyclic Redundancy Check (CRC) 10.
10 MODBUS PROTOCOL DESCRIPTION AND SET-UP ALTOSONIC V12 10.4.1 Address Field (Device Address) The address field of a message frame contains: ASCII mode 2 characters RTU mode 8 bits Valid slave addresses are 1...247. Address 0 is used for a broadcast to address all slaves. 10.4.2 Function Field The function field of a message frame contains: ASCII mode 2 characters RTU mode 8 bits Valid function codes are 1...127. The function code tells the slave which kind of action to perform.
ALTOSONIC V12 MODBUS PROTOCOL DESCRIPTION AND SET-UP 10 The contents of the error checking field for the entire message depend on the transmission mode. Error check in ASCII mode transmission For detecting errors in the entire message the error-checking field contains two ASCII characters. The error check characters are the result of a Longitudinal Redundancy Check (LRC) calculation.
10 MODBUS PROTOCOL DESCRIPTION AND SET-UP ALTOSONIC V12 • Input Registers: data, for example originating from electrical inputs of the slave or results from calculations in the slave, can be stored in "input registers". By nature the data in this area is "readonly": the master has only access to read this data. • Holding Registers: the master has access to this area to read the data but as well to set or change the value of data (write).
MODBUS PROTOCOL DESCRIPTION AND SET-UP 10 ALTOSONIC V12 10.5.3 Function 03: READ HOLDING REGISTERS Function 03 reads the contents of 1 to 125 contiguous holding registers in the slave. The maximum number of registers at each request is limited to 125 16-bit registers: 125 integers, 62 long integers, 62 floats, 31 doubles or 31 long longs. Request The request message specifies the starting register and the quantity of registers to be read. Registers are addressed starting from zero. Registers 1...
10 MODBUS PROTOCOL DESCRIPTION AND SET-UP ALTOSONIC V12 10.5.5 Function 05: WRITE SINGLE COIL Function 05 writes the status of a logical (Boolean or ON/OFF) variable. This function is not used, as in this application Boolean variables are not used as individual entities. Boolean variables are represented by means of specific bits packed in 32 bit data word (type “Long”). 10.5.6 Function 06: WRITE SINGLE HOLDING REGISTER Function 06 presets a value into a single holding register.
MODBUS PROTOCOL DESCRIPTION AND SET-UP 10 ALTOSONIC V12 Header Slave address Function Sub-function Data Hi+Lo Error check Trailer -- 11(h) 00 00(h) A1B8(h) -- -- 08(h) Table 10-6: Diagnostics All sub-functions are supported. 10.5.8 Function 15: WRITE MULTIPLE COILS Function 15 writes the status of 1...2000 contiguous logical (Boolean or ON/OFF) variables. This function is not used, as in this application Boolean variables are not used as individual entities.
10 MODBUS PROTOCOL DESCRIPTION AND SET-UP Header Slave address Function Starting address -- 11(h) 10(h) Hi 9C(h) Quantity of points Low 41(h) Hi 00(h) Low 02(h) ALTOSONIC V12 Error check Trailer -- -- Table 10-8: Response multiple holding registers If the request is not applicable, an exception response will be sent. For more information, refer to Exception responses on page 120. 10.5.
ALTOSONIC V12 MODBUS PROTOCOL DESCRIPTION AND SET-UP 10 Exception codes (for example) Code Name Meaning 01 Illegal function The function code in the request is not an allowable action for the slave. 02 Illegal data address The data address received in the request is not an allowable address for the slave. Table 10-10: Exception codes 10.6 Handling of large data types The standard Modbus specification does not explain how data types larger than 16 bits should be handled.
10 MODBUS PROTOCOL DESCRIPTION AND SET-UP Data Type Address range ALTOSONIC V12 Numbers of registers to request for each data type Modicon compatible Not Modicon compatible Integer (16 bit) 3000...3999 1 1 Long integer (32 bit) 5000...5999 2 1 Double (64 bit) 6000...6999 4 1 Float (32 bit) 7000...7999 2 1 Long long (64 bit) 8000...
ALTOSONIC V12 MODBUS PROTOCOL DESCRIPTION AND SET-UP 10 10.6.3 Single precision floating point (32 bit), transmit sequence Single precision floating-point numbers are stored in 32-bit registers, represented using the IEEE 754 encoding. In IEEE 754-2008 the 32-bit base 2 format is officially referred to as binary32. binary32 It was called single in IEEE 754-1985.
10 MODBUS PROTOCOL DESCRIPTION AND SET-UP ALTOSONIC V12 10.6.4 Double precision floating point (64 bit), transmit sequence Double precision floating-point numbers are stored in 64-bit registers, represented using the IEEE 754 encoding. In IEEE 754-2008 the 64-bit base 2 format is officially referred to as binary64. It was called double in IEEE 754-1985.
ALTOSONIC V12 MODBUS PROTOCOL DESCRIPTION AND SET-UP 10 Doubles transmit order IEEE 754 (1) 40h (2) 10h (3) 80h (4) 00h (5) 00h (6) 20h (7) 00h (8) 00h Normal mode (1) 40h (2) 10h (3) 80h (4) 00h (5) 00h (6) 20h (7) 00h (8) 00h Reversed mode (3) 80h (4) 00h (1) 40h (2) 10h (7) 00h (8) 00h (5) 00h (6) 20h 10.6.5 Long long (64 bit integer), transmit sequence Example 64 bit integer value 4.616.330.355.545.210.880 (= 4010 8000 0020 0000 hexadecimal).
10 MODBUS PROTOCOL DESCRIPTION AND SET-UP ALTOSONIC V12 10.7 Default settings By means of a number of parameters the Modbus communication link can be adjusted to one’s needs or preferences.
ALTOSONIC V12 MODBUS PROTOCOL DESCRIPTION AND SET-UP 10 10.8 Modbus register mapping Registers are mapped to specific address ranges according to both data and register type: Data type Register type Read command(s) Write command(s) Address range Integer (16 bit) Input register 4 n.a. 3000..3499 Holding register 3 6, 16 3500..3999 Long integer (32 bit) Input register 4 n.a. 5000..5499 Holding register 3 6, 16 5500..5999 Double (64 bit) Input register 4 n.a. 6000..
10 MODBUS PROTOCOL DESCRIPTION AND SET-UP ALTOSONIC V12 10.8.3 Input Registers (read-only): Long integer (32-bit); address range 5000-5499 Relative address Absolute address non-Modicon Modicon-comp. 0 5000 5000-5001 (2 regs) Name Parameter / Variable (short description) TestRegister uint 32 Test Register INFORMATION! Register reserved for testing communications and protocol handling with this type of register, without affecting the operation of the flowmeter. 128 www.krohne.
MODBUS PROTOCOL DESCRIPTION AND SET-UP 10 ALTOSONIC V12 Alarm / status events Relative address Absolute address Name non-Modicon Modicon-comp. Register Bit 1 5001 5002-5003 (2 regs) Alarm& Status Pipe A FailAll (bit 0) Parameter / Variable (short description) Down all channels, no custody transfer standard (FAILED) Alarm flag indicating none of the acoustic paths is operational.
10 MODBUS PROTOCOL DESCRIPTION AND SET-UP Name ALTOSONIC V12 Relative address Absolute address non-Modicon Modicon-comp. Register Bit Parameter / Variable (short description) 1 5001 5002-5003 (2 regs) Alarm& Status Pipe S Reset Totals (bit 18) Reset occurred: All Totalisers Status signal indicating totalisers have been reset. Note: this operation is only allowed for specially authorized personnel.
MODBUS PROTOCOL DESCRIPTION AND SET-UP 10 ALTOSONIC V12 Relative address Absolute address Name non-Modicon Modicon-comp. Register Bit Parameter / Variable (short description) 4 5004 5008-5009 (2 regs) Alarm& Status Channel_3 A Ch_Signal_ Lost (bit 3) Channel 3 Down: Signal lost Alarm flag indicating path failure (channel 3): signal lost.
10 MODBUS PROTOCOL DESCRIPTION AND SET-UP ALTOSONIC V12 Relative address Absolute address Name non-Modicon Modicon-comp. Register Bit Parameter / Variable (short description) 7 5007 5014-5015 (2 regs) Alarm& Status Channel_6 A Ch_Signal_ Lost (bit 3) Channel 6 Down: Signal lost Alarm flag indicating path failure (channel 6): signal lost.
ALTOSONIC V12 MODBUS PROTOCOL DESCRIPTION AND SET-UP 10 INFORMATION! Register reserved for testing communications and protocol handling with this type of register, without affecting the operation of the flowmeter. Relative address Absolute address non-Modicon Modicon-comp.
10 MODBUS PROTOCOL DESCRIPTION AND SET-UP ALTOSONIC V12 INFORMATION! Register reserved for testing communications and protocol handling with this type of register, without affecting the operation of the flowmeter. Transit Time Measuring Process Values Relative address Absolute address non-Modicon Modicon-comp.
ALTOSONIC V12 MODBUS PROTOCOL DESCRIPTION AND SET-UP 10 Measured / entered process parameters Relative address Absolute address non-Modicon Modicon-comp.
10 MODBUS PROTOCOL DESCRIPTION AND SET-UP ALTOSONIC V12 Statistics / diagnostics variables Averages and standard deviations are calculated for a number of flow variables. The number of measurement values that are included in this calculation are specified by means of a preset parameter. Averages and standard deviations are evaluated and updated each time a number of measurements equal to this parameter has been collected.
ALTOSONIC V12 MODBUS PROTOCOL DESCRIPTION AND SET-UP 10 Relative address Absolute address non-Modicon Modicon-comp. 43-48 7043-7048 (6 regs) 7086-7097 (12 regs) Name Parameter / Variable (short description) Ch_Reliab Channel reliability, channel 1-6 Reliability per channel (in %); array with values for channel 1 through 6.
10 MODBUS PROTOCOL DESCRIPTION AND SET-UP ALTOSONIC V12 10.8.8 Holding Registers (read/write): Float (32-bit) floating-point, address range 75007999 Relative address Absolute address non-Modicon Modicon-comp. 0 7500 7500-7501 (2 regs) Name Parameter / Variable (short description) TestRegister float32 Test Register Register reserved for testing communications and protocol handling with this type of register, without affecting the operation of the flowmeter. 138 www.krohne.
ALTOSONIC V12 MODBUS PROTOCOL DESCRIPTION AND SET-UP 10 Flowmeter configuration Relative address Absolute address non-Modicon Modicon-comp. 1 7501 2 7502 Name Parameter / Variable (short description) 7502-7503 (2 regs) Diameter Internal diameter 7504-7505 (2 regs) MeterConstat_Fwd Internal diameter of meter body at the measuring section (in m). Meter constant forward Meter constant, forward flow direction.
10 MODBUS PROTOCOL DESCRIPTION AND SET-UP Relative address Absolute address non-Modicon Modicon-comp. 21 7521 7542-7543 (2 regs) ALTOSONIC V12 Name Parameter / Variable (short description) Rev_LowFlowCutoff Low Flow cutoff reverse For reverse flow rates with smaller magnitude than this cutoff value, totalisers stop counting and outputs (such as the frequency output) will be disabled; flow rate indication will still be available (in m/s).
ALTOSONIC V12 MODBUS PROTOCOL DESCRIPTION AND SET-UP 10 Viscosity configuration Relative address Absolute address non-Modicon Modicon-comp. 26 7526 7552-7553 (2 regs) Name Parameter / Variable (short description) VD_Reference_Visc Dynamic Viscosity at Reference Temperature Input value of dynamic viscosity (in cP), used as base value to correct reference viscosity to actual viscosity at process temperature and pressure.
10 MODBUS PROTOCOL DESCRIPTION AND SET-UP ALTOSONIC V12 Totalisers Relative address Absolute address non-Modicon Modicon-comp. 1 8001 8004-8007 (4 regs) Name Parameter / Variable (short description) Fwd_Proces Totaliser PROCESS Forward Accumulated volume (with normal working meter) in forward flow direction, at line conditions (in ml (=10-6 m3)).
NOTES 11 ALTOSONIC V12 04/2013 - 4002643502 - MA ALTOSONIC V12 R02 en www.krohne.
© KROHNE 04/2013 - 4002643502 - MA ALTOSONIC V12 R02 en - Subject to change without notice. KROHNE product overview • • • • • • • • • • • • Electromagnetic flowmeters Variable area flowmeters Ultrasonic flowmeters Mass flowmeters Vortex flowmeters Flow controllers Level meters Temperature meters Pressure meters Analysis products Products and systems for the oil & gas industry Measuring systems for the marine industry Head Office KROHNE Messtechnik GmbH Ludwig-Krohne-Str. 5 47058 Duisburg (Germany) Tel.