8 SERVICE
88
ALTOSONIC V12
www.krohne.com 04/2013 - 4002643502 - MA ALTOSONIC V12 R02 en
Step 6: Mount the retraction tool
Fasten the lower part of the retraction tool (the ball valve) on the transducer adapter by means
of 4 bolts M 8 x 130 (grade 12.9) using the 6 mm hexagon screw driver. Check the O-ring
sealing, if damaged replace (size 40.87 x 3.53 mm).
Tighten the M8 bolts with appropriate force using the 6 mm hexagon angled key.
Make sure that the ball valve is in fully open position.
Measure the position of the transducer adapter inside the transducer port (depth in the center,
touching the bronze nut) with the ruler.
Memorize and note this value as "Depth".
INFORMATION!
When properly tightened there should still be some clearance between both plates of the
adapter.
Figure 8-24: Mount the ball valve of the retraction tool to the flowmeter
1 Transducer, fastening nut and transducer adapter inside the flowmeter.
2 Wedge shaped adapter plate.
3 Ball valve of the retraction tool
4 Use a 6 mm hexagon key to tighten the four M8 x 130 bolts.
INFORMATION!
In the figure above the square wedge shaped adapter plate is shown. This can also be the round
adapter plate, dependent from the type of flowmeter.
Figure 8-25: Measure depth
1 Bronze fixation nut.
2 Insert ruler up to bronze fixation nut.
3 Read the depth.