Manual

SERVICE 8
95
ALTOSONIC V12
www.krohne.com04/2013 - 4002643502 - MA ALTOSONIC V12 R02 en
Loosen all coax cables on the bottom of the electronic unit (the number of cables depend on
the version of the flowmeter).
Slide the new electronics unit partially back into the housing.
Push the metal pullers back to their original position.
Don't use excessive force, otherwise the connector at the backside can be damaged!
Screw the electronics unit back to the housing.
Re-install the display and make sure not to kink the display's flat ribbon cable.
Replace cover and tighten by hand.
Connect power.
CAUTION!
Please pay attention that the same amount of force is applied on both pullers, otherwise the
connector at the backside can be damaged.
Figure 8-36: Location of connectors
1 Position of connectors for transducer cables
2 Metal pullers
DANGER!
Electrostatic discharge (ESD) can damage electronic parts. Make sure to discharge yourself by
wearing a wrist strap. If no wrist strap is available, ground yourself by touching a metal surface
that is grounded.