Manual

Installation and operating instructions 18 BATCHFLUX
4.2 Operating and check
elements
The following operating elements are featured
on the hand-held communicator HHT 010:
HHT 010
+ –
Totalizer
I P
Overrange
Flow-
rate
190.47
m3/hr
M M M M M M
Display, 1st line
Display, 2nd line
Display, 3rd line: arrows to
identify
display
Flowrate current flowrate
Totalizer + totalizer
totalizer
sum totalizer (+ and –)
Overrange I overranging, current output
I
P overranging, pulse output P
Keys for operator control of signal
converter
Compass field, signals actuation of a key
4.3 Function of the keys
The cursor, flashing part of display, has a
gre
y background in the following descriptions.
To start operator control
Measuring mode
Operator
control mode
PLEASE NOTE: when “YES“ is set
under Fct. 3.4 ENTRY CODE,
“CodE 1 - - - - - - - - - appears in the
display after pressing key
.
Now enter the 9-keystroke Entry Code 1:
(each keystroke acknowledged by “ * “).
To terminate o
perator control
Press key repeatedly until one of the
following menus Fct. 1.0 OPERATION, Fct.
2.0 TEST
or Fct. 3.0 INSTALL. is displayed.
Press key
Store new parameters:
Confirm by pressing key
. Measuring mode
continued with the new parameters.
Do not store new
parameters:
Press key
to display “STORE.NO“.
Measuring mode continued with the “old“
parameters after pressing key
.
S T O R E Y E S
F c t . 3. 0
I N S T A L L.
1 3 . 5 7 1
m
3
/ h r
F c t . 1. 0
O P E R A T I O N
4