Installation and operating instructions 38 BATCHFLUX
7.2 Test of measuring range Q, Fct. 2.1
Please note!
Do not open the housing of the
BATCHFLUX IFM 5015 K.
Danger of contamination with substances
likely to destroy the moisture barrier of the
electronic equipment (e.g. if CIP or SIP
cleaned from the outside).
Therefore, please contact your KROHNE
Service engineer before you open the
housing.
Switch off
power source
before o
pening the housing
For this test a measured value can be
simulated in the range of -110 to +110
percent of Q
100%
(for set full-scale range
see Fct. 1.1 FULL SCALE).
Power the system.
Press the following keys for this test:
Keys Displayed Description
If “YES“ is set under Fct. 3.4 ENTRY CODE, key-in the
9-keystroke CODE 1:
Fct. 1.0 OPERATION
Fct. 2.0 TEST
Fct. 2.1 TEST Q
SURE NO
SURE YES
0 PERCENT Current, pulse and status outputs indicate the
corresponding values.
± 10 PERCENT
or
± 50 PERCENT
Select with keys and
± 100 PERCENT
± 110 PERCENT
Fct. 2.1 TEST Q End of test; actual measured values again present at
outputs.
(2x) 3x
- - - - - - - - - - - - / - - - Measuring mode
7