Instruction Manual

BM 102 45
Function (Fct.) Input Range Description
2.2.0 TESTS
2.2.1 TEST OUTPUT Select 4 mA, 20
mA, Other
This function allows the current
output to be tested.
The output can be set to one of the
listed values. With a reference
ammeter, the calibration of the
current output can be verified.
2.3.0 SERVICE Restricted access factory
configuration menu. These
parameters may be accessed via
Fct. 3.2 “Specialist PSW
(password).
3.0 ACCESS RIGHTS
Yes or no. Enter 9-
character code if
“Yes”.
3.1 MAINTENANCE PSW
No
Disables the access lock on the
configuration menu.
The password must contain exactly 9
characters.
E, R or U are used only. The
password is displayed in a scrambled
format. It allows KROHNE to decode
the password in case it was forgotten.
3.2 SPECIALIST PSW See KROHNE
Service centre or
BM 102 Service
Manual for code.
Specialist access to the Service
Parameters.
4.0 WATCH STATUS Displays the status of the device.
5.0 HART
®
VARIABLES
5.1 MANUFACTURER
KROHNE
Read only menu.
5.2 MODEL
BM 102
Read only menu.
5.3 FLD DEV REV
1.0
Field device revision.
Read only menu.
5.4 SOFTWARE REV
1.0
Software revision.
Read only menu.
5.5 HARDWARE REV
1.0
Hardware revision.
Read only menu.
5.6 DEVICE ID Read only menu. The device ID is
also the serial number of the device.
5.7 MESSAGE 32 bytes of ASCII characters
5.8 DESCRIPTOR 16 bytes of ASCII characters.
Equivalent to PCSTAR2 Fct. 1.4.8
5.9 DATE Month Day Year (xx / xx / xx).
5.10 NUM RESP PREAM Number of preamble in the response
frame of the device.
5.11 TAG Tag name of the BM 102.
5.12 POLL ADDRESS Address of the device.