User Manual

17
ALLOWABLE TIME AT HIGH TMP.
300G/ 800G METER
0
10
20
30
40
50
60
70
80
90
100
110
120
120 130 140 150 160 170 180 190 200
TEMP. C
TIME MIN.
Initial
Initial
Initial
2. Electrical installation
2.1 Location and connecting cables
Location
Do not expose the compact flow meter to direct sunlight. Install a sunshade if necessary.
Connecting cables
To conform to protection category requirements, observe the following recommendations:
Fit blanking plug PG 16 and apply sealant to unused cable entries.
Do not kink cables directly at cable entries.
Provide water drip point (U bend in cable).
Do not connect rigid conduit to cable entries.
If cables are a tight fit, enlarge inside diameter of cable gland by removing the appropriate
ring(s) from the seal.