Fct. No. Text Description and settings
3.9.0 TUBE PARAMS Submenu 3.9.0 primary head specific parameters
Password protected, see Fct. 3.8.8
3.9.1 CF1 Density coefficient 1
Input the value stamped on the primary head’s data plate
or perform on site calibration as described in Sect. 5.12.
3.9.2 CF2 Density coefficient 2
Input the value stamped on the primary head’s data plate
or perform on site calibration as described in Sect. 5.12.
3.9.3 CF3 * Reference strain
Displays the value stamped on the primary head’s data
plate.
3.9.4 CF4 * Reference temperature
Displays the value stamped on the primary head’s data
plate.
3.9.5 CF5 *Primary head constant
Displays the value stamped on the primary head’s data
plate.
3.9.6 DSS CF6 Density strain slope
Displays the value which is mentioned on the cal. Sheet
3.9.7 DTS CF 7 Density temperature slope
Displays the value which is mentioned on the cal. Sheet
3.9.8 FSS CF8 Slope strain
Displays the value which is mentioned on the cal. Sheet
3.9.9 FTS CF9 Slope temperature
Displays the value which is mentioned on the cal. Sheet
3.9.10 D.REF.HIGH Density adjustment, High Point
3.9.11 D.REF.LOW Density adjustment, Low Point
3.10.0 CONC.MEAS Concentration measurement option when installed
3.10.1 SOLUTE R20 See separate Concentration instruction manual
3.10.2 SOLUTE K1 See separate Concentration instruction manual
3.10.3 SOLUTE K2 See separate Concentration instruction manual
3.10.4 LIQUID See separate Concentration instruction manual
3.10.5 LIQUIDR20 See separate Concentration instruction manual
3.10.6 LIQUID K1 See separate Concentration instruction manual
3.10.7 LIQUID K2 See separate Concentration instruction manual
3.11.0 SERIAL I/O RS485 or Modbus option, if installed
3.11.1 PROTOCOL See separate RS485 or Modbus instruction manual
3.11.2 ADDRESS See separate RS485 or Modbus instruction manual
3.11.3 BAUDRATE As per 3.11.1
* Only accessible if allowed through password in Fct. 3.8.8.