Fct. No. Text Description and settings
4.0 PARAM.ERROR Main Menu 4.0 Parameter error
4.1 Not Used
4.2.0 CUR.OUTP.I Range settings incorrect
LOW SCALE ≥ FULL SCALE
4.2.1 LOW SCALE Low scale range for current output I see Fct. 3.3.3
4.2.2 FULL SCALE Full scale range for current output I see Fct. 3.3.4
4.3.0 ZERO Zero calibration incorrect.
The measured zero offset must be less than ±10 %
of the primary head’s full scale flow rating.
4.3.1 ZERO SET Zero calibration see Fct. 3.1.7
4.3.2 PRIMRY.TYPE Type of primary head see Fct. 3.1.5
4.4 Not Used
4.5 Not Used
4.6.0 PULS.OUTP. P Range setting incorrect
LOW SCALE ≥ FULL SCALE
4.6.1 LOW SCALE Low scale range for pulse output see Fct. 3.4.3
4.6.2 FULL SCALE Full scale range for pulse output see Fct. 3.4.4
4.7.0 PROC. ALARM Minimum and maximum limits incorrect
MIN.LIMIT > 96 % of MAX.LIMIT
4.7.1 MIN.LIMIT Minimum limit for range checking see Fct. 3.5.3
4.7.2 MAX.LIMIT Maximum limit for range checking see Fct. 3.5.4
4.8.0 SYS.CTRL.S Minimum and maximum limits incorrect
MIN.LIMIT > 96 % of MAX.LIMIT
4.8.1 MIN. LIMIT Minimum limit for condition checking see Fct. 3.7.3
4.8.2 MAX. LIMIT Maximum limit for condition checking see Fct. 3.7.4
4.5 Reset / Quit Menu - Totalizer reset and status indication acknowledgement
Totalizer reset
Button Display Description
10.36
kg
Measurement mode
↵ CodE 2
––
Enter access Code 2 for reset/quit menu: ↑ →
↑ → RESET.TOTAL Totalizer reset menu
Only appears if ”yes” programmed in Fct. 3.8.5. Reset
enable No or Yes. If ”no” is programmed ”status light”
only appears. See next section.
→ RESET.YES If the reset function is enabled RESET YES will be
shown, press ↵ to execute the function.
To cancel the reset operation press ↑ to get RESET NO
and then press ↵
If the reset function is disabled by menus Fct. 3.8.5 or
3.8.6 then BLOCKED is displayed. Press ↵ to continue
↵↵ 0.00
kg
Assuming RESET YES was selected the totalisers will
now be cleared.