User Manual
42
View status message(s) and quit
Button Display Description
0.36
kg/min
∇
Measurement mode
The presence of the ∇ marker above Status on the
display indicates the presence of warning messages in
the status list.
↵ CodeE 2
– –
∇
Enter access code for reset/quit menu: ↑ →
↑→ RESET.TOTAL
∇
Totalizer reset menu.
↑ STATUS.LIST
∇
View/Quit Status message menu
→ ≡ 1 Err ≡
MASS FLOW
∇
This display shows that there is just 1 warning in the list,
in this case MASS FLOW. The ≡ symbols indicate that
this is a new error and not one that has been previously
acknowledged. Use either the ↑ or → keys to view other
messages in the list. Otherwise press ↵ to exit.
→ ≡ 1 Err ≡
QUIT YES
∇
At the end of the message list the QUIT YES prompt is
shown. Selecting YES will clear if possible messages in
the list.
To cancel the operation press ↑ to get QUIT NO and
then press ↵
↵ STATUS.LIST Assuming the conditions that caused the message have
passed (i.e. mass flow is back within the meter’s range)
then the Status marker, ∇ will disappear.
↵ 0.36
kg/min
Assuming RESET YES was selected, the totalisers will
now be cleared.