User Manual
60
H
Max. Limit
Min. Limit
H
Hysteresis, H = 2% Max Limit
INACTIVE
ACTIVE
FLOW
Process alarm characteristics
For example, a process requires that the process fluid temperature stays within the range 30 to
40°C and the low level signal is required to indicate that the temperature has gone out of range.
Begin from measuring mode
Return to measuring mode.
NOTE:
For functions other than range checking Fct. 3.5.3. and 3.5.4. are not accessible.
Key Display
line 1 line 2
→↑↑ Fct. (3).0. INSTALL
→4x↑ Fct. 3.(5).0. ALARM.OUT.A
→ Fct. 3.5.(1). FUNCTION A
→ (OFF)
↑ (MASS FLOW)
: (MASS TOTAL)
: (DENSITY)
↑ (TEMPERATURE)
↵ Fct. 3.5.(1). FUNCTION A
↑ Fct. 3.5.(2). ACTIV.LEVEL
→ (ACTIVE.HIGH)
↑ (ACTIVE LOW)
↵ Fct. 3.5.(2). ACTIV.LEVEL
↑ Fct. 3.5.(3). MIN. LIMIT
→ Input minimum temperature
↵ Fct. 3.5.(3). MIN. LIMIT
↑ Fct. 3.5.(4). MAX. LIMIT
→ Input maximum temperature
↵ Fct. 3.5.(4). MAX. LIMIT
4x↵