The default value of the custody transfer protection code when leaving the factory is
↵→↑↵↑→↵→↑
This password may be changed using Menu 3.8.7.. However, this can only be done if custody
transfer protection is disabled first, as described above.
Begin from measuring mode.
If the two entered passwords are different the message "CODE WRONG" is displayed.
The message has to be acknowledged with the key and input must be repeated with Fct.
3.8.7.. Then the protection state 'active' or 'inactive' can be chosen with Fct.3.8.6..
NOTE:
If the input of the custody transfer password is wrong, a 9-character code is displayed. With this
character code the manufacturer can decode the password in cases where the password is
lost.
It is also possible to protect just the mass totaliser. Menu 3.8.5. ENABL.RESET determines
whether the operator can clear the mass total from the ACKNOWLEDGE/RESET MENU.
Begin from measuring mode.
Key Display
line 1 line 2
Fct. (1).0 OPERATOR
2x Fct. (3).0 INSTALL.
Fct. 3.(1).0 BASE PARAM.
7x Fct. 3.(8).0 USER DATA
Fct. 3.8.(1) LANGUAGE
6x Fct. 3.8.(7) CODE 3
CodeE 3 - - - - - - - -
Enter new custody transfer
password two times.
Fct. 3.8.(7) CODE 3
4x
Key Display
line 1 line 2
Fct. (1).0 OPERATOR
2x Fct. (3).0 INSTALL
7x Fct. 3.(8).0 USER DATA
4x Fct. 3.8.(5) ENABL.RESET
(YES)
(NO)
Fct. 3.8.(5) ENABL.RESET
4x +110.25 kg
Mass Total Display
CodE 2 - -
↑→ RESET MASS
BLOCKED
Totaliser Reset is disabled
↵↵