4 APPLICATION PROFILE
30
IFC 300
www.krohne.com 02/2011 - 4001086601 - AD IFC 300 PROFIBUS R01 en
4.3.3 DIAGNOSIS_EXTENSION (if "Classic Status" and "Diagnosis" selected)
Octet number
Bit number
Subparameter
Comment*
1 0 ADC_ELEC_TEMPERATURE (S) electronic temperature
1 ADC_GAIN_ERROR (S) gain error
2 ADC_F_FREQ_TOO_HIGH (S) field frequency too high
3 Reserved
4 ADC_COIL_TEMPERATURE (S) coil temperature
5 ADC_F_CURR_DEVIATION (S) field current deviation
6 ADC_F_COIL_BRIDGED (S) field coil bridged
7 ADC_F_COIL_BROKEN (S) field coil broken
2 0 ADC_ELECTRODE_NOISE (S) electrode noise
1 ADC_FLOW_PROFILE (S) flow profile
2 ADC_LINEARITY (S) linearity
3 ADC_ELECTRODE_SYM (S) electrode symmetry
4 Reserved
5 Reserved
6 ADC_PIPE_NOT_FULL (S) pipe not full
7 ADC_EMPTY_PIPE_0 (S) empty pipe (Uncertain)
3 0 Reserved
1 ADC_FIELD_FREQ_HIGH (F) field frequency too high
2 ADC_EMPTY_PIPE_1 (F) empty pipe (Application Error)
3 ADC_DC_OFFSET (F) DC Offset
4 Reserved
5 ADC_FIELD_CURR_LOCAL (F) field current local
(F) field current local(F) field current local
(F) field current local
6 ADC_SENSOR_LOCAL (F) sensor local
(F) sensor local(F) sensor local
(F) sensor local
7 ADC_SENSOR_GLOBAL (F) sensor global
(F) sensor global(F) sensor global
(F) sensor global
4 0 ADC_FLOW_EXCEED_LIM0 (F) flow exceeding limit
1 ADC_FLOW_EXCEED_LIM1 (F) flow exceeding limit
2 MEAS_AVAILABLE_NO (I) measurement value not available
3 POWER ON ERROR (I) power fail
4 ADC_TEST_SENSOR (C) test sensor
5 ADC_WARNING_W1 (S) uncertain measurement
6 ADC_WARNING_W2 (F) application error
7 ADC_SENSOR_ELECTRONIC (F) sensor electronic
(F) sensor electronic(F) sensor electronic
(F) sensor electronic