User's Manual

6 OPERATION
30
OPTIWAVE 1010
www.krohne.com 09/2014 - 4003537401 - MA OPTIWAVE 1010 R01 en
6.1 Software configuration
6.1.1 General notes
How to configure the software for device communication.
Before the program can send data to and receive data from the device, it is necessary to add
elements to a project structure. The project structure is built in the PACTware Project pane.
The Project pane is on the left side of the window.
This procedure will open the communication port, but does not start the communication with the
device.
6.1.2 Procedure
HOST PC is shown in the Project pane. Go to the main toolbar and click on the View menu
button. Click on "Device catalog F3" to open the Device Catalog pane.
Double click on HART Communication in the Device Catalog pane. The "COM4" element is
added below "HOST PC in the project structure.
Click on "OK" to save changes or "Cancel" to cancel the new configuration.
Double click on the "OPTIWAVE 1010" item in the Device Catalog pane. This step will add the
device DTM to the project structure in the Project pane.
i The software is correctly configured for device communication, but the port is not open and
the device cannot communicate with the software at this time.
(a) Double click on the "OPTIWAVE 1010" element in the project structure (Project pane), or (b)
Right click on "OPTIWAVE 1010" element in the project structure (Project pane) and select
"Parameter" in the drop-down list box.
i This step will open the OPTIWAVE 1010 Parameter (configuration) window.
Right click on the "OPTIWAVE 1010" element in the Project pane and select "Connect" in the
drop-down list box.
i The software is correctly configured for device communication and the port is open, but the
device cannot communicate with the software at this time. End of the procedure.
INFORMATION!
This procedure will open the communication port, but DOES NOT start the communication with
the device.