User's Manual

8 TECHNICAL DATA
50
OPTIWAVE 1010
www.krohne.com 09/2014 - 4003537401 - MA OPTIWAVE 1010 R01 en
Process connections
Welded version
Pipe sizes Stainless steel (1.4404 / 316L): Ø42.4×2; Ø60.3×2; Ø60.3×2.77; Ø60.3×3.9
Hastelloy® C-22® (2.4602) : Ø42.4×2; Ø60.3×2; Ø60.3×2.77; Ø60.3×3.9
Flange version
EN 1092-1 DN4080 (Form B1, B2, C, D, E or F) in PN16 / 40; others are available on request
ASME B16.5 1½...3¨ (RF or FF) in 150 lb / 300 lb; others are available on request
JIS 40...80A in 10K
Electrical connections
Power supply 14.530 VDC; min./max. value for an output of 22 mA at the terminal
Maximum current 22 mA
Current output load R
L
[Ω] ((U
ext
-14.5 V)/22 mA). For more data, refer to
Minimum power supply
voltage
on page 53.
Cable entry Standard: M20×1.5; Option: ½ NPT
Cable gland Standard: none
Options: M20×1.5 (cable diameter: 6...10 mm / 0.2...0.39¨); others are available on
request
Cable entry capacity (terminal) 0.52.5 mm²
Input and output
Current output / HART®
Output signal
420 mA HART
®
or 3.820.5 mA acc. to NAMUR NE 43 4
Resolution ±3 µA
Temperature drift Typically 50 ppm/K (150 ppm/K maximum)
Digital temperature drift Max. ±15 mm / 0.6¨ for the full temperature range
Error signal High: 22 mA; Low: 3.6 mA acc. to NAMUR NE 43 5
Approvals and certification
CE This device fulfils the statutory requirements of the EC directives. The
manufacturer certifies successful testing of the product by applying the CE mark.
Vibration resistance EN 60068-2-6 / IEC 61298-3
10-82.2 Hz: 0.15 mm; 82.2-1000 Hz: 20 m/s²