OPTIWAVE 7400-24 C
www.krohne.com12/2015 - 4004901901 - MA OPTIWAVE7400-24 R01 fr
3.7.5 Standpipes (stilling wells and bypass chambers)
Use a standpipe if:
There is highly conductive foam in the tank.
The liquid is very turbulent or agitated.
There are too many other objects in the tank.
The device is measuring a liquid (petro-chemicals) in a tank with a floating roof.
The device is installed in a horizontal cylindrical tank (refer to the end of this section)
Stilling wells - general notes
Installation in tanks containing one liquid and foam
Drill an air circulation hole (max. Ø10 mm / 0.4¨) in the stilling well above the maximum level.
Remove the burr from the hole.
Installation in tanks containing one liquid or more without foam
Drill an air circulation hole (max. Ø10 mm / 0.4¨) in the stilling well above the maximum level.
Figure 3-12: Installation recommendations for standpipes (stilling wells and bypass chambers)
1 A stilling well solution
2 A bypass chamber solution
3 Air circulation hole
4 Level of the liquid
CAUTION!
The standpipe must be electrically conductive.
The inside diameter of the standpipe must not be more than 5 mm / 0.2
¨
over the diameter of
the antenna (for a high-dielectric constant liquid).
The standpipe must be straight. There must be no sudden changes in internal diameter
greater than 1 mm / 0.04
¨
.
The standpipe must be vertical.
Recommended surface roughness: <
±
0.1 mm / 0.004
¨
.
Make sure that there are no deposits at the bottom of the standpipe.
Make sure that there is liquid in the standpipe.