User Manual

OPERATION
6
67
OPTIWAVE 7400-24 C
www.krohne.com12/2015 - 4004901901 - MA OPTIWAVE7400-24 R01 fr
3. Service menu
4. Master menu
6.4 Further information on device configuration
6.4.1 Quick Setup (Commissioning)
Use this procedure to change the tank height, tank type, output function, output range and give
the top and bottom measuring limits. Values and parameters that can be changed are shown
between the « ... » marks in the illustrations that follow. Push the keypad buttons in the correct
sequence:
Procedure
Menu
No.
Function Function description Selection list Default
3.0.0 SERVICE Advanced settings. The settings in this
menu are protected with a password.
Only approved personnel can change
the parameters in this menu. For more
data, speak or write to your local sales
office.
Menu
No.
Function Function description Selection list Default
4.0.0 MASTER Factory settings. The settings in this
menu are protected with a password.
Only approved personnel can change
the parameters in this menu. For more
data, speak or write to your local sales
office.
CAUTION!
Make sure that you do this procedure before you use the device. The settings in this procedure
have an effect on the performance of the device.
Screen Steps Description
[>
>>
>], [
] and [>
>>
>]. Default screen.
Enter configuration mode (2.0.0
SUPERVISOR).
[>
>>
>], [^
^^
^], [
], [
], [>
>>
>] and [^
^^
^]. Enter the password (the default password is
shown). If it is necessary to change the
password, menu item 2.7.5 PASSWORD.
[>
>>
>] Push this button to start the quick set-up
procedure.