OPERATION
74
OPTIWAVE 7400-24 C
www.krohne.com 12/2015 - 4004901901 - MA OPTIWAVE7400-24 R01 fr
Repeat the last 3 steps to complete the table.
Push [^
^^
^] to go back to the "STORE" screen.
Push [
] or [
] to set the screen to STORE YES
STORE YESSTORE YES
STORE YES and push [^
^^
^].
i The device will store the data for the strapping table and go back to normal mode.
The device will give more accurate volume readings if you give more conversion data in these
areas:
Surfaces with curves.
Sudden changes in the cross section.
Refer also to the illustration that follows:
How to delete a volume or mass table
Enter the supervisor menu.
Push 7 × [
], [>
>>
>], and [
] to go to 2.8.2 DELETE TABLE.
Push [>
>>
>] and [
] to set the parameter to YES
YESYES
YES.
Push [^
^^
^] to go back to the "STORE" screen.
Push [
] or [
] to set the screen to STORE YES
STORE YESSTORE YES
STORE YES and push [^
^^
^].
i The device will delete the data for the strapping table and go back to normal mode. The
"CONVERSION" and "ULLAGE CONV." data are not available in normal mode.
Figure 6-6: A plot of points for a volume or mass table
1
2