User Manual

TECHNICAL DATA
8
89
OPTIWAVE 7400-24 C
www.krohne.com12/2015 - 4004901901 - MA OPTIWAVE7400-24 R01 fr
Display LCD display
128 × 64 pixels in 8-step greyscale with 4-button keypad
Interface languages 4 language pack options (the language is given in the customer order):
1 English, French German and Italian
2 English, French,Spanish and Portuguese
3 English, Chinese (simplified), Japanese and Russian
4 English, Czech, Polish and Turkish
Measuring accuracy
Resolution 1mm/ 0.04¨
Repeatability ±1mm/ ±0.04¨
Accuracy Standard: ±2mm/ ±0.8¨, when distance <10m/ 33ft;
±0.2% of measured distance, when distance > 10 m / 33 ft
Reference conditions acc. to EN 61298-1
Reference conditions acc. to EN 61298-1Reference conditions acc. to EN 61298-1
Reference conditions acc. to EN 61298-1
Temperature +15...+25°C / +59...+77°F
Pressure 1013 mbara ±50 mbar / 14.69 psia ±0.73 psi
Relative air humidity 60% ±15%
Target Metal plate in an anechoic chamber
Operating conditions
Temperature
TemperatureTemperature
Temperature
Ambient temperature -40+80°C/ -40+176°F
Ex: see supplementary operating instructions or approval certificates
Storage temperature -50+85°C/ -58+185°F
Process connection temperature
(higher temperature on request)
Metallic Horn (machined) / Metallic Horn (sheet metal) antenna:
Metallic Horn (machined) / Metallic Horn (sheet metal) antenna:Metallic Horn (machined) / Metallic Horn (sheet metal) antenna:
Metallic Horn (machined) / Metallic Horn (sheet metal) antenna:
Standard: -50+150°C/ -58+302°F
Option: -50+200°C/ -58+392°F
(the process connection temperature must agree with the temperature limits of the
gasket material. Refer to "Materials" in this table.)
Ex: see supplementary operating instructions or approval certificates
Drop antenna (PTFE):
Drop antenna (PTFE):Drop antenna (PTFE):
Drop antenna (PTFE):
-50+150°C/ -58+302°F (the process connection temperature must agree with
the temperature limits of the gasket material. Refer to "Materials" in this table.)
Ex: see supplementary operating instructions or approval certificates
Drop antenna (PP):
Drop antenna (PP):Drop antenna (PP):
Drop antenna (PP):
-40+100°C/ -40+212°F (the process connection temperature must agree with
the temperature limits of the gasket material. Refer to "Materials" in this table.)
Ex: see supplementary operating instructions or approval certificates
Drop antenna (PEEK):
Drop antenna (PEEK):Drop antenna (PEEK):
Drop antenna (PEEK):
-20+200°C/ -4+392°F (the process connection temperature must agree with
the temperature limits of the gasket material. Refer to "Materials" in this table.)
Ex: see supplementary operating instructions or approval certificates
Hygienic antenna (PEEK):
Hygienic antenna (PEEK):Hygienic antenna (PEEK):
Hygienic antenna (PEEK):
-20+150°C/ -4+302°F (the process connection temperature must agree with
the temperature limits of the gasket material. Refer to "Materials" in this table.)
Ex: see supplementary operating instructions or approval certificates
Pressure
PressurePressure
Pressure