Stilling wells - general notes
Installation in tanks containing one liquid and foam
Drill an air circulation hole (max. Ø10 mm / 0.4¨) in the stilling well above the maximum level.
Remove the burr from the hole.
Installation in tanks containing one liquid or more without foam
Drill an air circulation hole (max. Ø10 mm / 0.4¨) in the stilling well above the maximum level.
Drill 1 or more liquid circulation holes in the stilling well (if there is more than 1 liquid in the
tank).
i These holes help the liquid to move freely between the stilling well and the tank.
Remove the burr from the hole.
Stilling wells: floating roofs
If the device must be installed on a tank with a floating roof, install it in a stilling well.
Figure 3-12: Installation recommendations for standpipes (stilling wells and bypass chambers)
1 A stilling well solution
2 A bypass chamber solution
3 Air circulation hole
4 Level of the liquid
CAUTION!
The standpipe must be electrically conductive.
The inside diameter of the standpipe must not be more than 5 mm / 0.2
¨
over the diameter of
the antenna (for a high-dielectric constant liquid).
The standpipe must be straight. There must be no sudden changes in internal diameter
greater than 1 mm / 0.04
¨
.
The standpipe must be vertical.
Recommended surface roughness: <
±
0.1 mm / 0.004
¨
.
Make sure that there are no deposits at the bottom of the standpipe.
Make sure that there is liquid in the standpipe.