User Manual

OPERATION
6
69
OPTIWAVE 7400-24 C
www.krohne.com12/2015 - 4003518001 - MA OPTIWAVE7400-24-R01-en
3. Service menu
4. Master menu
2.7.7 If it is not easy to read the data on the LCD
display screen, this menu item operates the
display screen light. If you set this menu
item to YES
YESYES
YES, the light comes on.
YES, NO NO
2.8.0 CONV. TABLE
2.8.1 INPUT TABLE The device uses a conversion table
(strapping table) to convert measurements
to volume, mass and flow rate readings. The
readings are shown in normal mode. Give
the number of entries on the table (min. 2;
max. 30). Select an entry (01...30) and enter
the level and the related volume / mass /
flow rate value for that entry. Push [^
^^
^] to
confirm the entry values. Continue the
procedure until the device has data for all
the entries. For more data, refer to
How to
configure the device to measure volume or
mass
on page 76.
min. 2 entries
max. 30 entries
(level / volume, mass or
flow rate)
0 entries
2.8.2 DELETE TABLE This menu item erases the data in the
conversion table.
YES, NO NO
2.9.0 CONFIG/RESET
2.9.1 SAVE This menu item is not available. YES, NO NO
2.9.2 RECALL This menu item is not available. YES, NO NO
2.9.3 RESTART This menu item starts the device again. YES, NO NO
2.9.4 RESET FACT. If you set this menu item to "YES", the
device goes back to its initial settings (set by
the manufacturer in the factory).
YES, NO NO
1 This value is given in the customer order data
2 Units and range depend on the output function, length unit and volume unit selected
3 If the device has the LCD display option, this depends on data given in the customer order
Menu
No.
Function Function description Selection list Default
3.0.0 SERVICE Advanced settings. The settings in this
menu are protected with a password.
Only approved personnel can change
the parameters in this menu. For more
data, speak or write to your local sales
office.
Menu
No.
Function Function description Selection list Default
4.0.0 MASTER Factory settings. The settings in this
menu are protected with a password.
Only approved personnel can change
the parameters in this menu. For more
data, speak or write to your local sales
office.
Menu
No.
Function Function description Selection list Default