Owner's manual

2/2
Connection diagrams for outputs 2.3
Milliammeter
Totalizer
– electronic (EC)
– electromechanical (EMC)
1 2
3 4
I = 0/4 - 20 mA
U
ext
15...20 V DC 20...32 V DC
R
i
0...500 250...750
U
ext
15 V DC
I = 0/4 - 20 mA
R
i
500
Stromausgang
I
active
Stromausgang
I
passive
6
5
Pulse output P
passive
for electronic (EC) or
electromechanical (EMC) totalizers
U
ext
32 V DC/ 24 V AC
I
max
150 mA
(incl. status output)
U
int
15 V DC from current output
Operation with current output:
I
max
3 mA
Operation without current output:
I
max
23 mA
R
15 V
I
max
U
ext
U
ext
DC voltage,
external power source (U
ext
),
note connection polarity
External voltage source (U
ext
),
DC or AC voltage,
connection polarity arbitrary
Active mode
The current output supplies the power for
operation of the outputs.
Passive mode
External power source required for operation of
the outputs.
I
max
Status output S
passive
U
ext
32 V DC/ 24 V AC
I
max
150 mA
(incl. pulse output)
Pulse output P
active (and current output I
active
)
for electronic (EC) totalizers
with and without current output I
Status output S
active
with and without current output I
U
int
15 V DC
from current output
I
max
3 mA
Operation with
current output
I
max
23 mA
Operation without
current output
Load
Load