Owner's manual

4/6
Fct. Text Description and settings
1.04 DISPLAY Display functions
DISP.FLOW Select flow display
NO DISP.
user unit, factory set is “Liter/hr” or “US MGal/day (see Fct. 3.05)
m3/hr PERCENT
Liter/Sec BARGRAPH (value and bargraph display in %)
US.Gal/min
Press
key to transfer to subfunction “DISP. TOTAL.”.
DISP.TOTAL. Select totalizer display
NO DISP. (totalizer switched on but not displayed)
OFF (totalizer switched off)
• +TOTAL. • –TOTAL. • +/–TOTAL. • SUM (Σ)
• ALL (displaysingle counts or all)
m3 Liter US.Gal
user unit, factory set is “Liter” or “US MGal” (see Fct. 3.05).
Press
key to transfer to format setting.
Format setting
• Auto (exponent notation)
# . ####### ##### . ###
## . ###### ###### . ##
### . ##### ####### . #
#### . #### ########
Press
key to transfer to subfunction “DISP.MSG”.
DISP.MSG. Additional messages required in measuring mode?
NO YES (cyclic change with displays of measured values)
Press
key to return to Fct. 1.04 DISPLAY.
1.05 CURRENT I Current output I
FUNCT. I Select function for current output I
OFF (switched off)
1 DIR. (1 flow direction)
2 DIR. (forward/reverse flow, F/R flow measurement)
Press key
, transfer to subfunction
“RANGE I”;
if
“2 DIR.”
selected, transfer to subfunction
“REV.RANGE”!
RANGE I Select measuring range
0 - 20 mA 4 - 20 mA (fixed ranges)
Press key
to transfer to subfunction
“I ERROR“.
I ERROR Select error value
0 mA 3.6 mA (only with range 4-20 mA) 22 mA
Press key
to revert to
Fct. 1.05 CURRENT. I.