Owner's manual
4/7
Fct. Text Description and settings
1.06 PULS.OUTP. P Pulse output P
→
FUNCTION P Select function for pulse output P
• OFF (switched off)
• 1 DIR. (1 flow direction)
• 2 DIR. (forward/reverse flow, F/R measurement)
Press
↵
key to transfer to subfunction “SELECT P “.
→
SELECT P Select pulse type
• 100 Hz • PULSE/VOL. (pulses per unit volume, flow rate)
• 1000 Hz • PULSE/TIME (pulses per unit time for 100% flowrate)
Press
↵
key to transfer to subfunction “PULSWIDTH“.
When 100 Hz and 1000 Hz selected, return to
Fct. 1.06 PULS.OUTP. P, pulse width 50% cyclic.
→
PULSWIDTH Select pulse width
• 50 mSec • 100 mSec • 200 mSec • 500 mSec • 1 Sec
Press
↵
key to transfer to subfunction “VALUE P“.
→
VALUE P Set pulse value per unit volume (appears only when
“PULSE/VOL.“ has been set under “SELECT P“).
• xxxx PulS/m3 • xxxx PulS/Liter • xxxx PulS/US.Gal
• xxxx PulS/ user unit, factory set is “Liter“ or “US MGal“ (see Fct. 3.05).
Setting range „xxxx“ is dependent on the pulse width and the
full-scale range: P
min
= F
min
/ Q
100%
P
max
= F
max
/ Q
100%
Press
↵
key to return to Fct. 1.06 “PULS.OUTP. P“.
→
VALUE P Set pulse value per unit time (appears only
when „PULSE/TIME“ has been set under „SELECT P“).
• xxxx PulS/Sec (=Hz) • xxxx PulS/min • xxxx PulS/hr
• xxxx PulS/user unit, factory set is “hr“ or “day“ (see Fct. 3.05)
“xxxx“ setting range is dependent on pulse width, see above.
Press
↵
key to return to Fct. 1.06 “PULS.OUTP. P“.
1.07 IND. OUTP. S Status output S
• ALL ERROR • FATAL ERROR • OFF • ON
• F/R INDIC. (F/R indication for forward/reverse measurement)
• TRIP. POINT Setting range: 002 - 115 PERCENT
• EMPTY PIPE (appears only when this option is installed)
(
Press
↵
key to transfer to number setting.)
Press
↵
key to return to Fct. 1.07 “IND.OUTP. S“.
Fct. Text Description and settings
2.00 TEST Test menu
2.01 TEST Q Test measuring range Q
Precautionar
y query
• SURE NO
Press
↵
key to return to Fct. 2.01 “TEST Q”.
• SURE YES
Press
↵
key, then use
↑
or
↓
key to
select value:
-110 / -100 / -50 / -10 / 0 / +10 / +50 / +100 / +110 PCT.
of set full-scale range Q
100%
.
Displayed value present at outputs I and P.
Press
↵
key to return to Fct. 2.01 “TEST Q”.
2.02 HARDW. INFO Hardware information and error status
Before consulting factory, please note down all 6 codes.
→ MODUL ADC X . X X X X X . X X
Y Y Y Y Y Y Y Y Y Y
Press
↵
key to transfer to
“MODUL IO“.
→ MODUL IO X . X X X X X . X X
Y Y Y Y Y Y Y Y Y Y
Press
↵
key to transfer to
“MODUL DISP.“.
→ MODUL DISP. X . X X X X X . X X
Y Y Y Y Y Y Y Y Y Y
Press
↵
key to return to Fct. 2.02
“HARDW. INFO“.