Fct. Text Description and settings
3.00 INSTALL. Installation menu
3.01 LANGUAGE Select language for display texts
GB / USA (English) F (French)
D (German) others on request
Press
key to return to Fct. 3.01
“LANGUAGE”.
3.02 FLOWMETER Set data for primary head
DIAMETER Select size from meter size table
DN 10 - 1000 mm equivalent to 3/8 - 40 inch
Select with
or
key.
Press
key to transfer to subfunction “FULL SCALE”.
FULL SCALE Full-scale range for flow Q
100%
To set, refer to Fct. 1.01 “FULL SCALE” above.
Press
key to transfer to subfunction “GKL VALUE”.
VALUE P Pulse value (Fct. 1.06
“VALUE P”
) has been changed.
With the “old” pulse values the output frequency (F)
would have been exceeded or not reached.
P
min
= F
min
/ Q
100%
P
max
= F
max
/ Q
100%
Check new values!
GKL VALUE Set primary constant GKL
see primary head nameplate.
Ran
ge: 1.0000 - 9.9999
Press
key to transfer to subfunction
“FIELD. FREQ.”.
FIELD FREQ. Magnetic field frequency
Values: 1/6 or 1/18
of power frequency, see nameplate.
Press
key to transfer to subfunction “FLOW DIR.”;
DC units only: to transfer to subfunction
“LINE FREQ.”.
LINE FREQ. Normal line frequency in your country
Please note: This function is only provided for units
with DC power supply to suppress line-frequency interference.
Values:
50 Hz
and
60 Hz
Press
key to transfer to subfunction “FLOW DIR.”.
FLOW DIR. Define flow direction (in F/R mode: forward flow).
Set according to direction of arrow on primary head:
+ DIR. – DIR.
Select using
or
key.
Press
key to return to Fct. 3.02 “FLOWMETER”.
3.03 ZERO SET Zero calibration
Note: Carry out only at “0” flow and with completely filled
measuring tube!
Precautionar
y query
CALIB. NO
Press
key to return to Fct. 3.03 “ZERO SET”.
CALIB. YES Press
key to start calibration.
Duration approx. 1590 seconds, current flowrate
displayed in the selected unit (see Fct. 1.04 “DISP. FLOW”).
A “WARNING” sign appears when flowrate “>
0
“;
acknowledge by pressing
key.
STORE NO (do not store new zero value)
STORE YES (store new zero value)
Press
key to return to Fct. 3.03 “ZERO SET”.
3.04 ENTRY CODE Entry code required to enter setting mode?
NO (= entry with only)
YES (= entry with and Code 1: → → → ↵ ↵ ↵ ↑ ↑ ↑ )
Press
to return to Fct. 3.04 “ENTRY CODE”.