Owner's manual

4/10
Reset totalizer and cancel error messages, RESET / QUIT menu 4.6
Cancel error messages in RESET / QUIT menu
Key Display Description
- - - - - - - - - - - - / - - - Measuring mode
CodE 2 - - Key in entry code 2 for RESET / QUIT
menu: ↑→
↑ → ERROR QUIT. Menu for error acknowledgement
QUIT. NO Do not delete error messages,
press twice = return to measuring mode
QUIT. YES Delete error messages
ERROR QUIT. Error messages deleted
- - - - - - - - - - - - / - - - Return to measuring mode
Reset totalizer(s) in RESET / QUIT menu
Key Display Description
- - - - - - - - - - - - / - - - Measuring mode
CodE 2 - - Key in entry code 2 for RESET / QUIT
menu: ↑→
↑ → ERROR QUIT. Menu for error acknowledgement
TOTAL. RESET Menu for resetting totalizer
RESET NO Do not reset totalizer,
press twice = return to measuring mode
RESET. YES Reset totalizer
RESET QUIT. Totalizer reset
- - - - - - - - - - - - / - - - Return to measuring mode
Examples of setting the signal converter 4.7
The cursor, flashing part of display, is shown below in bold type.
Change measuring range of current output and value for error messages (Fct. 1.05):
Change measuring range from 04-20 mA to 00-20 mA
Change value for error messages from 0 mA to 22 mA
Key Display Description
If “YES” set under Fct. 3.04 ENTRY CODE, key in the
9-keystroke CODE 1 now: → → → ↵ ↵ ↵ ↑ ↑ ↑
Fct. 1.00 OPERATION
Fct. 1.01 FULL SCALE
4x Fct. 1.05 CURRENT I
FUNCT. I
→↵ RANGE I If “REV. RANGE” appears here,
press keys and again.
04-20 mA Old current range
00-20 mA New current range
I ERROR
0 mA Old value for error messages
2x 22 mA New value for error messages
Fct. 1.05 CURRENT I
Fct. 1.00 OPERATION
STORE YES
- - - - - - - - - - - - / - - - Measuring range with new data for the current output