Owner's manual
4/10
Reset totalizer and cancel error messages, RESET / QUIT menu 4.6
Cancel error messages in RESET / QUIT menu
Key Display Description
- - - - - - - - - - - - / - - - Measuring mode
↵ CodE 2 - - Key in entry code 2 for RESET / QUIT
menu: ↑→
↑ → ERROR QUIT. Menu for error acknowledgement
→ QUIT. NO Do not delete error messages,
press ↵ twice = return to measuring mode
↑ QUIT. YES Delete error messages
↵ ERROR QUIT. Error messages deleted
↵ - - - - - - - - - - - - / - - - Return to measuring mode
Reset totalizer(s) in RESET / QUIT menu
Key Display Description
- - - - - - - - - - - - / - - - Measuring mode
↵ CodE 2 - - Key in entry code 2 for RESET / QUIT
menu: ↑→
↑ → ERROR QUIT. Menu for error acknowledgement
↑ TOTAL. RESET Menu for resetting totalizer
→ RESET NO Do not reset totalizer,
press ↵ twice = return to measuring mode
↑ RESET. YES Reset totalizer
↵ RESET QUIT. Totalizer reset
↵ - - - - - - - - - - - - / - - - Return to measuring mode
Examples of setting the signal converter 4.7
The cursor, flashing part of display, is shown below in bold type.
• Change measuring range of current output and value for error messages (Fct. 1.05):
• Change measuring range from 04-20 mA to 00-20 mA
• Change value for error messages from 0 mA to 22 mA
Key Display Description
→ If “YES” set under Fct. 3.04 ENTRY CODE, key in the
9-keystroke CODE 1 now: → → → ↵ ↵ ↵ ↑ ↑ ↑
Fct. 1.00 OPERATION
→ Fct. 1.01 FULL SCALE
4x ↑ Fct. 1.05 CURRENT I
→ FUNCT. I
→↵ RANGE I If “REV. RANGE” appears here,
press keys → and ↵ again.
→ 04-20 mA Old current range
↑ 00-20 mA New current range
↵ I ERROR
→ 0 mA Old value for error messages
2x ↑ 22 mA New value for error messages
↵ Fct. 1.05 CURRENT I
↵ Fct. 1.00 OPERATION
↵ STORE YES
↵ - - - - - - - - - - - - / - - - Measuring range with new data for the current output