7.4 Faults and symptoms during start-up and process flow measurement
Before contacting Krohne Service, please read through the
following tables. THANK YOU !
Most faults and symptoms occurring with the compact flowmeters can be eliminated
with the aid of the following tables.
For greater clarity, faults and symptoms in the tables are divided into 2 parts and
various groups.
Part 1
Signal converter IFC 010 B (B = basic version), without display
and without HHT or CONFIG user program (see Sect. 6.1)
Groups: LED LED display (status messages)
I Current output
P Pulse output
LED / I / P LED display, current output and pulse output
Part 2
Signal converter IFC 010 D (D = display version) and
Signal converter IFC 010 B (B = basic version), without display
but with CONFIG user program (see Sect. 6.1)
Groups: D Display
I Current output
P Pulse output
S Status indication output
D / I / P / S LED display, current output,
status output
Part 1 Converter IFC 010 B (B = basic version), without display
and without HHT or CONFIG operator program
Group LED Fault / Symptom Cause Remedial action
LED 1 LED flashes red/green Overranging of A/D converter, Reduce flowrate; if unsuccessful,
current or pulse output test as described in Sect. 7.5.
Measuring tube drained, Fill measuring tube.
A/D converter overranged.
LED 2 LED flashes red Fatal Error, hardware and/or Replace signal converter
software error (see Sect. 8.4) or contact
Krohne Service.
LED 3 cyclic flashing of red LED, Hardware fault, Watchdog Replace signal converter
approx 1 second activated. (see Sect. 8.4) or contact
Krohne Service.
LED 4 LED shows red continuously Hardware fault Replace signal converter
(see Sect. 8.4) or contact
Krohne Service.