Error limits at reference conditions 10.2
Pulse output
± F Error in % of flowrate (actual value):
Curve A: DN 10 – 600 /
3
/8”- 24”
v 0.4 m/s or v 1.3 ft/s : ± 0.5 % of measured value
v 0.4 m/s or v 1.3 ft/s : ± 0.002 m/s or 0.0066 ft/s
Curve B: DN 700 – 1000 / 28” – 40”
v 0.25 m/s or v 0.8 ft/s : ± 0.8 % of measured value
v 0.25 m/s or v 0.8 ft/s : ± 0.002 m/s or 0.0066 ft/s
Q Actual flowrate
Q
F
Flow for error limit vF = 0.25 or 0.8 ft/s (see Flow tables)
v Flow velocity in m/s and ft/s
v
F
Flow velocity in m/s and ft/s at Q
F
(see Flow tables)
Reference conditions
Product Water, 10 to 30°C / 50 to 86°F
Electrical conductivity > 300 µS/cm (µmho/cm)
Power supply (line voltage) U
N
(± 2%)
Ambient temperature 20 to 22°C / 68 to 71.6°F
Warm-up time 30 minutes
Straight inlet run > 10
x
DN
(DN = meter size)
Straight outlet run > 3
x
DN
Primary heads properly grounded and centered
Current output same as above error limit for pulse output plus ...
0 to 20 mA:
± 0.05 %
of full-scale range in each case
4 to 20 mA:
± 0.062 %
}
}
± F 3
%
2
1
0.8
0.5
0.25 0.4 0.5 0.75 1 10 11 12
(0.8) (1.3) (1.6) (2.4) (3.2) (33) (36) (40)
Q
F
QF
B
A
v
ms
(ft /sec)
}