1.3 Electrical connection of separate primary head (F Versions)
1.3.1 General information on signal cable A and field current cable C
Use of the Krohne signal cable A with foil screen and magnetic shield will ensure proper operation
of the equipment.
Signal cable to be solidly laid.
Connect shields via stranded drain wires.
Underwater and underground installation possible.
Insulating material is flame-retardant to IEC IEC 332.1 / VDE 0742.
Signal cables are low in halogen, unplasticized, and stay flexible at low temperatures.
Si
gnal cable A (Type DS), with double shielding
1 Stranded drain wire, 1st shield, 1.5 mm
2
or AWG14
2 Insulation
3 Conductor 0.5 mm
2
or AWG 20 (3.1 red / 3.2 white)
4 Special foil, 1st shield
5 Inner sheath
6 Mu-metal foil, 2nd shield
7 Stranded drain wire, 2nd shield, 0.5 mm
2
or AWG20
8 Outer sheath
Field current cable C
with single shielding
Cross-section is dependent on required length of cable, see Table in Sect. 1.3.4.
1.3.2 Grounding of primary head
All flowmeters must be properly grounded.
The grounding conductor should not transmit any interference voltages.
Do not ground any other electrical device together with this conductor.
The primary head is connected to ground by means of an FE functional ground conductor.
Special information on grounding various primary heads is contained in the separate
installation instructions for primary heads.
These instructions also contain detailed descriptions on how to use grounding rings and
how to install primary heads in metal or plastic pipes or internally coated pipelines.