User Manual

Installation and operating instructions IFC 020
4/6
Fct. Text Description and settings
1.04 DISPLAY Display functions
DISP.FLOW Select flow display
NO DISP.
user unit, factory set is “Liter/hr” or “US MGal/day (see Fct. 3.05)
m3/hr PERCENT
Liter/Sec BARGRAPH (value and bargraph display in %)
US.Gal/min
Press key to transfer to subfunction “DISP. TOTAL.”.
DISP.TOTAL. Select totalizer display
NO DISP. (totalizer switched on but not displayed)
OFF (totalizer switched off)
+TOTAL TOTAL ±TOTAL SUM (Σ)
ALL (single totalizer or all)
Press key to transfer to display unit
m3 Liter US.Gal
user unit, factory set is “Liter” or “US MGal” (see Fct. 3.05).
Press key to transfer to format setting.
Format settin
g
• Auto (exponent notation)
# . ####### ##### . ###
## . ###### ###### . ##
### . ##### ####### . #
#### . #### ########
Press key to transfer to subfunction “DISP.MSG”.
DISP.MSG. Additional messages required in measuring mode?
NO YES (cyclic change with displays of measured values)
Press key to return to Fct. 1.04 DISPLAY.
1.05 CURRENT I Current output I
FUNCT. I Select function for current output I
OFF (switched off)
1 DIR. (1 flow direction)
2 DIR. (forward/reverse flow, F/R flow measurement)
Press key , transfer to subfunction “RANGE I”;
RANGE I Select measuring range
0 - 20 mA 4 - 20 mA (fixed ranges)
mA (user-defined range) I
0%
- I
100%
(Value I
0%
< I
100%
!) 0 - 16 mA 4 - 20 mA
To transfer to number setting, press key
!
Press key to transfer to subfunction “I ERROR“.
I ERROR Select error value
22 mA 0.0 to I
0%
mA (variable, see above if I
0%
> 1 mA)
To transfer to number setting, press key
!
Press key to revert to Fct. 1.05 CURRENT. I.