User Manual

Fct. Text Description and settings
1.06 PULS B1 Pulse output P
FUNCT. P Select function for pulse output P
OFF (switched off)
1 DIR. (1 flow direction)
2 DIR. (forward/reverse flow, F/R flow measurement)
Press key to transfer to subfunction “SELECT P”.
SELECT P Select pulse type
100 HZ PULSE/VOL. (pulses per unit volume, flowrate)
1000 HZ PULSE/TIME (pulses per unit time for 100% flowrate)
Press key to transfer to subfunction “PULSWIDTH”.
With selection 100 HZ and 1000 HZ,
return to Fct.1.06 PULSOUTPUT P (pulswidth 50% cyclic).
PULSWIDTH Select pulse width
50 msec 100 msec 200 msec 500 msec 1 sec
Press key to transfer to subfunction “VALUE P”.
VALUE P Set pulse value per unit volume (appears only when
“PULSE/VOL.” set under “SELECT P” above)
xxxx PulS/m3 xxxx PulS/Liter xxxx PulS/US.Gal
xxxx PulS/
user-defined unit, factory-set is “Liter” or “US M.Gal” (see Fct. 3.05)
Setting range
xxxx” is dependent on the pulse width and the
full-scale range: P
min
= F
min
/ Q
100%
P
max
= F
max
/ Q
100%
Press key to return to Fct.1.06
PULS. OUTPUT P”.
VALUE P Set pulse value per unit time (appears only when
PULSE/TIME”
set under
SELECT P” above)
xxxx PulSe/Sec (=Hz) xxxx PulSe/min xxxx PulSe/hr
xxxx PulSe/user-defined unit, factory-set is “hr” (see Fct. 3.05)
Setting range
xxxx” is dependent on the pulse width, see above
Press key to return to Fct.1.06
PULS. OUTPUT P””.
1.07 STATUS. S Status output S
ALL ERROR FATAL ERROR OFF ON
F/R INDIC. (F/R indication for forward/reverse measurement)
TRIP. POINT Settin
g range: 002 - 115 PERCENT
Press key to transfer to number setting
Press key to return to Fct. 1.07 “STATUS. S“.
1.08 CONTROL E Control input E (IFC 020 E only)
(IFC 020 E only) OFF (switched off)
OUTP. ZERO (set outputs to “min.values”)
TOTAL. RESET (reset totalizers)
ERROR. RESET(delete error messages)
Press key to return Fct. 1.08 “Control E”.
Installation and operating instructions IFC 020
Fct. Text Description and settings
2.00 TEST Testmenu
2.01 TEST Q Test measuring range Q
Precautionary query
SURE NO Press key to return to Fct. 2.01 “TEST Q”.
SURE YES Press key, then use or key to
select value: -110 / -100 / -50 / -10 / 0 / +10 / +50 / +100 / +110 PCT.
of set full-scale range Q
100%
.
Displayed value present at outputs I and P.
Press key to return to Fct. 2.01 “TEST Q”.
2.02 HARDW. INFO Hardware information and error status
Before consulting factory, please note down all 6 codes.
MODUL ADC X . X X X X X . X X
Y Y Y Y Y Y Y Y Y Y Press key to transfer to “MODUL IO“.
MODUL IO X . X X X X X . X X
Y Y Y Y Y Y Y Y Y Y Press key to transfer to “MODUL DISP.“.
MODUL DISP. X . X X X X X . X X Press key to transfer to “MODUL RS
Y Y Y Y Y Y Y Y Y Y
MODUL RS X . X X X X X . X X Press key to return to Fct. 2.02
Y Y Y Y Y Y Y Y Y Y “HARDW. INFO“.
4/7