Installation and operating instructions IFC 020
4/12
Examples of setting the signal converter 4.7
As an example the cursor, flashing part of display, is shown below in bold type.
• Change measuring range of current output and value for error messages (Fct. 1.05):
• Change measuring range from 04-20 mA to 00-20 mA
• Change value for error messages from 0 mA to 22 mA
Key Display Description
→ If “YES” set under Fct. 3.04 ENTRY CODE, key in the
9-keystroke CODE 1 now: → → → ↵ ↵ ↵ ↑ ↑ ↑
Fct. 1.00 OPERATION
→ Fct. 1.01 FULL SCALE
4x ↑ Fct. 1.05 CURRENT I
→ FUNCT. I
→↵ RANGE I
→ 04-20 mA Old current range
↑ 00-20 mA New current range
↵ I ERROR
→ 0 mA Old value for error messages
2x ↑ 22 mA New value for error messages
↵ Fct. 1.05 CURRENT I
↵ Fct. 1.00 OPERATION
↵ STORE YES
↵ - - - - - - - - - - - - / - - - Measuring range with new data for the current output