Low-flow cutoff 5.3
Fct. 1.03 L.F.CUTOFF
Press key.
Choice
OFF (fixed tripping point: ON = 0.1 % / OFF = 0.2 %
with 100Hz and 1000Hz, see Fct. 1.06; 1% resp. 2%)
PERCENT (variable tripping points: ON = 1 - 19 % / OFF = 2 - 20 %)
Select with or key.
Transfer to number setting using key (only if “PERCENT” selected).
1st number (cursor) flashes.
Settin
g the numerical value when “PERCENT” selected
01 to 19 (cutoff “on” value, left of hyphen)
02 to 20 (cutoff “off” value, right of hyphen)
Change flashing number (cursor) with the or key.
Shift cursor 1 place to right using key.
Press key to return to Fct. 1.03 L.F.CUTOFF.
Note: The cutoff “off” value must be greater than the cutoff “on” value.
Continuation see next page
Display 5.4
Fct. 1.04 DISPLAY
Press key.
DISP. FLOW = select unit for display of flowrate, press key
NO DISP. (no display)
m3/hr (cubic metres per hour)
Liter/Sec (litres per second)
US.Gal/min (US gallons per minute)
user-defined unit, factory-set: “Liter/hr” (litres per hour) or “US MGal/day”, see Sect. 5.14
PERCENT (percentage display)
BARGRAPH (numerical value and bar graph display in %)
Select with or key.
Press key to transfer to subfunction “DISP. TOTAL”.
DISP. TOTAL = select unit for totalizer display, press key
NO DISP. (no display)
OFF (internal totalizer switched off)
• + TOTAL. • – TOTAL. • +/– TOTAL. • SUM (
ΣΣ
) • ALL (sequential)
Select with or key.
Transfer to totalizer unit setting using key.
•m3 (cubic metres)
Liter (litres)
US.Gal (US gallons)
user-defined unit, factory-set: “Liter” or “US MGal”, see Sect. 5.14
Select with or key.
Transfer to totalizer format setting using key.