Current output I 5.6
Fct. 1.05 CURRENT I
Press → key.
→
FUNCT. I = Select function for current output, press → key
• OFF (switched off, no function)
• 1 DIR. (1 flow direction)
• 2 DIR. (2 flow directions, F/R mode, forward/reverse)
Select using ↑ or ↓ key.
Transfer to subfunction “RANGE I”
with ↵ key.
Exceptions: When “OFF” selected, return to Fct. 1.05 CURRENT I.
→
RANGE I = select measuring range, press → key
• 0 - 20 mA
fixed ranges
• 4 - 20 mA
• mA (user-defined value) I
0%
– I
100%
(value I
0%
< I
100%
!) 0-16 mA 4-20 mA
Press → key to transfer to number setting.
Select with ↑ or ↓ key.
Press key ↵ to transfer to subfunction “I
ERROR”.
→ I ERROR = set error value, press → key
• 22 mA (fixed value)
• 0.0 - I
0%
mA (variable value; only variable when I
0%
≥ 1 mA, see ”RANGE I“ above)
Select using ↑ or ↓ key.. Press → key to transfer to number setting.
Press key ↵ to return to Fct. 1.05 CURRENT I.
Please refer to Sect. 3.2 “Factory settings”.
Refer to Sect. 2.3 for connection diagrams, and to Sect. 5.14 for characteristics.
}