Installation and operating instructions IFC 020
5/9
5.11 Primary head
Please refer to Sect. 3.2 “factory settings”
Fct. 3.02 FLOW METER
Press key.
DIAMETER = set meter size (see instrument nameplate) press key
Select size from table of meter sizes:
DN 2.5 - 1000 mm equivalent to
1
/10 - 40 inch
Select using or key.
Transfer to subfunction “FULL SCALE”
with key.
FULL SCALE = set full-scale range, press key.
Set as described in Sect. 5.1.
Transfer to subfunction “GK V
ALUE” with key.
Note: if “VALUE P” is displayed after pressing key.
PULSE/VOL. is set under Fct. 1.06 PULS.OUTP. P, subfunction “SELECT P”. Because the
full-scale range Q
100%
has been changed, the output frequency (F) of the pulse output is over- or
undershot: P
min
= F
min
/ Q
100%
P
max
= F
max
/ Q
100%
Change pulse value accordingly, see Sect. 5.7 pulse output P, Fct. 1.06.
GK VALUE = set primary constant GK, press key.
1.0000 - 9.9999 (note information on instrument nameplate, do not change setting !)
Change flashing digit (cursor) with or key.
Shift cursor 1 place to right or left with or key.
Transfer to subfunction “FIELD
FREQ.” with key.
FIELD FREQ. = set magnetic field frequency, press key
1/2 • 1/18 (1/2, 1/6 1/18 or 1/36 of power frequency,
1/6 • 1/36 see instrument nameplate, do not change setting)
Select using or key.
Transfer to subfunction “FLOW DIR.”
with key.
(only for units with DC power supply, transfer to subfunction “LINE FREQ”).
LINE FREQ. = normal line frequency in your country, press key.
(Please note, applicable for flowmeters equipped with DC in power supply!)
50 Hz Select using the
or key.
60 Hz Transfer to subfunction “FLOW
DIR.” with
key.
FLOW DIR. = set flow direction, press key.
+ DIR. (for identification of flow direction, see “+” arrow on primary head;
- DIR. for F/R mode, identifies the “positive” flow direction)
Select using the or key.
Press key to return to Fct. 3.02 FLOW METER.
Zero check
, see Fct. 3.03 and Sect. 7.1.
}